Instruction manual

17
High Throughput Diffusion Pumps
DRAFT 12/23/04
Electrical Connections
Terminal connections for each pump are different depending on the source voltage available, the principle difference
being a Y or connection of the heaters. Table 13 provides the figure number and page of the wiring diagrams in this
manual. The specific wiring diagrams for each pump shows both Y and
connections and the source voltages for each
connection.
To wire the pump:
1. Check the heater for correct supply voltage and find
the appropriate wiring diagram. The correct voltage is
shown.
2. Check load balance by measuring the resistance of
each branch. The heater resistances are given on
their respective wiring diagrams.
3. Make the connections to the terminals in the
electrical main junction box at the foreline shown in
the appropriate outline drawing. Use flexible conduit
to make it easier to remove the pump for service.
4. Wire the thermal switches to a control mechanism to
ensure that power to the pump is shut off if one of the
switches opens. The thermal switch wires can be
found in the electrical box.
5. Complete the wiring of the pump and double-check
that the proper terminal arrangement has been used
for the source voltage at the site.
WARNING Improper operation, severe heater
damage, and danger to personnel
can result from an improperly wired
diffusion pump system.
WARNING High voltages (up to 480 V) can kill.
Always break the primary circuit to the
pump before starting work on the heater
or its wiring.
Table 13 Wiring Diagram Locations
Source Voltage HS-16 Wiring HS-20 Wiring HS-32 Wiring NHS-35 Wiring
200 Figure 13 on page 21 Figure 15 on page 23
240 Figure 11 on page 19 Figure 13 on page 21 Figure 16 on page 24 Figure 20 on page 28
280 Figure 11 on page 19
400 Figure 13 on page 21 Figure 20 on page 28
415 Figure 12 on page 20 Figure 14 on page 22 Figure 18 on page 26 Figure 21 on page 29
430 Figure 13 on page 21
440 Figure 20 on page 28
460 Figure 17 on page 25
480 Figure 13 on page 21 Figure 19 on page 27 Figure 22 on page 30
CAUTION Do not operate the pump at more than
5% over its rated voltage.