User`s manual

MENU SYSTEM : SELF TESTS
4
02.9676.0019 A 05-January-2004 59
Self Tests
Running Parameter Tests
The luminaire is capable of running self tests by using the TEST menu functions. For complete
details and values for each test refer to “Diagnostic Tests” on page 61.
Note: After 10 seconds of inactivity, the menu display will change to the default state showing the
address.
When running tests on multiple luminaires, a DMX termination connector is required at the last
luminaire in the link. (Refer to page 12 for more information regarding the construction of this
connector.)
Figure 4-2: Test Setup
To run tests:
Step 1. Press [Menu].
Step 2. Press [Up] / [Down] arrows until Te s t appears. Press [Enter].
Step 3. Use [Up] / [Down] arrows to select a parameter to test. Refer to “Menu System Functions”
on page 53 for a complete list of test parameters which can be chosen.
Step 4. Press [Enter] to run test.
Step 5. Press [Menu] to stop test at any time.
Note: Connected luminaires may not respond on the first action (may delay) of the master luminaire.
DATA THRU
ENTERMENU
AC PWR
DATA IN~
AC IN
DATA THRU
ENTERMENU
AC PWR
DATA IN~
AC IN
DATA THRU
ENTERMENU
AC PWR
DATA IN~
AC IN
Termination Connector
Data Thru Data Thru
Master luminaire containing
software version to download
Luminaire 1
requiring update
Luminaire 2
requiring update