User Manual
Table Of Contents
- 4343W X-ray Detector
- Table of Contents
- List of Tables
- List of Figures
- 1.0 Scope
- 2.0 Contraindication
- 3.0 Intended Use
- 4.0 Audience
- 5.0 Abbreviations
- 6.0 Definition of Symbols
- 7.0 Standards and Regulations
- 8.0 Description of the 4343W
- 8.1 Shipment Contents
- 8.2 Optional Parts
- 8.3 X-ray Detector Surfaces and Features
- 8.4 X-ray Detector Dimensions
- 8.5 X-ray Detector Specifications
- 8.6 Environmental Considerations
- 8.7 X-ray Detector Battery
- 8.8 Optional Varex Provided Inductive Battery Charger
- 8.9 Designing a Custom Inductive Battery Charger
- 8.10 Using a Third-Party Inductive Battery Charger
- 8.11 Power-on and Power-off Sequence
- 8.12 Reboot Sequence
- 8.13 LED Status Indicator Behavior
- 8.14 Position Sensing
- 8.15 Detector Sharing
- 9.0 The Detector Configuration File
- 10.0 System Software
- 11.0 Maintenance
- 12.0 Troubleshooting
- 13.0 Disposing of the X-ray Detector and Battery
- 14.0 Safety - Electromagnetic Interference
- 15.0 Regulatory
- Index
www.vareximaging.com 59
4343W X-ray Detector
REFERENCE MANUAL
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m 30 A/m Magnetic field should be that of a typical
location in a typical professional healthcare
environment.
IEC 61000-4-6 3 Vrms
150 kHz to 80 MHz
6V/m (in ISM bands
between 0.15MHz and
80MHz n)
80% AM (at 1kHz e)
N/A
Battery power
equipment not
connected to mains
-
IEC 61000-4-3 3 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
3 V/m
80 MHz to 2.7 GHz
Where P is the maximum power in W, d is
the minimum separation distance in m and E
is the Immunity Test Level in V/m.
If the X-ray Detector complies with
Immunity Test Levels for this test, the 30cm
minimum separation distance (in 5.2.1.1 f)
may be replaced with minimum separation
distances calculated from the higher
Immunity Test Levels.
Table 15 Test Specs for Enclosure Port Immunity to RF Wireless Communications Equipment
Test
Frequency
Band Service Modulation
Max
Power
Distance
Immunity
Test Level
385 380-390 TETRA 400 Pulse Modulation
b) 18 Hz
1.8 0.3 27
450 430-470 GMRS 460, FRS
460
FM c) +/- 5 kHz
deviation 1 kHz sine
20.3 28
710
745
780
704-787 LTE BAND 13,
17
Pulse modulation b)
217 Hz
0.2 0.3 9
810
870
930
800-960 GSM 1800;
TETRA 800;
iDEN 820;
CDMA 850;
LTE Band 5
Pulse modulation b) 18
Hz
0.3 28
1720
1845
1970
1700-1990 GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse modulation
b) 217 Hz
20.3 28
Table 14 ESD, Transient/Burst, Surge, Voltage Variation, Magnetic Fields (Continued)
Immunity test
IEC 60601-1-2
test level
Compliance Electromagnetic environment
E
6
d
---
P=