User Manual

Table Of Contents
www.vareximaging.com 41
4343W X-ray Detector
REFERENCE MANUAL
10.8.1 Acquiring Radiography Images
Radiography provides single-shot, high-resolution images for diagnosis.
1 Choose the desired imaging mode from the Mode Drop Down Menu.
Figure 39 Mode Drop Down Menu
2
Make sure Hardware Handshaking is selected, see Figure 34.
3 Click Acquire Image button or click Acquire Image in the Acquisition menu under the Menu
Bar to begin acquiring images.
Figure 40 Acquire Image Button
Note
The OEM workstation should have the ability to employ the PREP and
Expose_OK signals. See Section 10.3 or Section VSPADU-UTILITY in 03236
FP2032_VarexSmartPanel_SoftwareInterfaceSpecification.PDF for an
example of controlling the X-ray Generator.
Note
In order to acquire an image with X-ray exposure in vTrigger mode,
Hardware Handshaking must be enabled, see Figure 34, or an image
without exposure will be returned.