User Manual
Table Of Contents
- 4343W X-ray Detector
- Table of Contents
- List of Tables
- List of Figures
- 1.0 Scope
- 2.0 Contraindication
- 3.0 Intended Use
- 4.0 Audience
- 5.0 Abbreviations
- 6.0 Definition of Symbols
- 7.0 Standards and Regulations
- 8.0 Description of the 4343W
- 8.1 Shipment Contents
- 8.2 Optional Parts
- 8.3 X-ray Detector Surfaces and Features
- 8.4 X-ray Detector Dimensions
- 8.5 X-ray Detector Specifications
- 8.6 Environmental Considerations
- 8.7 X-ray Detector Battery
- 8.8 Optional Varex Provided Inductive Battery Charger
- 8.9 Designing a Custom Inductive Battery Charger
- 8.10 Using a Third-Party Inductive Battery Charger
- 8.11 Power-on and Power-off Sequence
- 8.12 Reboot Sequence
- 8.13 LED Status Indicator Behavior
- 8.14 Position Sensing
- 8.15 Detector Sharing
- 9.0 The Detector Configuration File
- 10.0 System Software
- 11.0 Maintenance
- 12.0 Troubleshooting
- 13.0 Disposing of the X-ray Detector and Battery
- 14.0 Safety - Electromagnetic Interference
- 15.0 Regulatory
- Index
www.vareximaging.com 39
4343W X-ray Detector
REFERENCE MANUAL
Figure 36 System Settings window
10.8 Image Acquisition
Acquisition can be performed after Offset and Gain Calibration have been performed. This
section describes how to acquire images.
Before performing acquisition, settings should be checked to ensure that the desired corrections
are applied to acquired frames. Table 11 lists the types of corrections that can be applied.
Table 11 Types of Corrections that can be applied during Acquisition
Type Images transferred to Workstation
Raw Exposed Frame and Post-Offset Frame
Corrected A single fully corrected frame
Offset Corrected A single offset corrected frame, without gain and defect
correction
Gain Corrected A single gain corrected frame, without defect correction
Note
When acquiring Corrected, Offset Corrected, and Gain corrected images; if
preview is enabled and additional preview frame will be delivered to the
workstation prior to any additional frames.