User Manual

Table Of Contents
www.vareximaging.com 11
4343W X-ray Detector
REFERENCE MANUAL
8.5 X-ray Detector Specifications
Table 6 X-ray Detector Specifications
Sensor
Detector Amorphous Silicon active TFT/PIN diode Technology
Scintillator CSI Premium, CSI Standard, and DRZ+
Pixel Matrix 3072 (v) x 3072 (h)
Pixel Pitch 139 µm
Active Area 3062 (v) x 3062 (h) DRZ+, 3052 (v) x 3052 (h) CSI
Electronics
Battery Lithium-ion
Battery Charger 1 or 3 Bay, Inductive
ADC 16-bit
Mechanical
Housing Plastic with Carbon Fiber entrance window
Weight (with Battery) DRZ+ 3.1 kg (6.83 lbs), CSI 3.3 kg (7.27 lbs)
Load Support 200 kg over diameter 40 mm at center, 300 kg entire surface
Surface Temperature Rated to not exceed 42°C
Wireless Communication
Signal Strength Requires > -70 dBm or no image will be acquired
Standard IEEE 802.11ac/a/n
Interface USB
Security
WEP
WPA
WPA2
Operating Voltage DC 5V
Radio
Antenna 2 x IPEX connector for 2T2R
Frequencies
UNII - 1: 5150MHz - 5250MHz
UNII - 3: 5725MHz - 5850MHz
Note: Subject to local regulations
Modulation
802.11a: OFDM (BPSK, QPSK, 16-QAM, 64-QAM)
802.11n: OFDM (BPSK, QPSK, 16-QAM, 64-QAM)
802.11ac: OFDM (BPSK, QPSK, 16-QAM, 64-QAM, 256-QAM)