Technical information
6
Setup preset Installation example Sensors measured/
calculated values
required parameters
Hydrostatic Tank Gauging (HTG)
HTG P1
L00-NRF590-14-00-08-yy-006
• Pressure sensor
(P1, bottom)
• level (calculated) • Tank Ref Height
•local gravity
• density of the measured medium
• Min HTG Level
(minimum level at which HTG
measurement is possible; should be
slightly above the position of the P1
sensor)
• P1 Position
• P3 Position (only for the "HTG P1,3"
mode)
HTG P1,3
!
Note!
This mode should be
used in non-
atmospheric tanks (e.g.
pressurised tanks)
L00-NRF590-14-00-08-yy-009
• Pressure sensor
(P1, bottom)
• Pressure sensor
(P3, top)
HTG P1,2
L00-NRF590-14-00-08-yy-007
• Pressure sensor
(P1, bottom)
• Pressure sensor
(P2, middle)
• level (calculated)
• density of the measured
medium (calculated)
•Tank Ref Height
•local gravity
• Min HTG Level
(minimum level at which HTG
measurement is possible; should be
slightly above the position of the P2
sensor)
• P1 Position
• P1-P2-Distance
• P3 Position (only for the "HTG
P1,2,3" mode)
HTG P1,2,3
!
Note!
This mode should be
used in non-
atmospheric tanks (e.g.
pressurised tanks)
L00-NRF590-14-00-08-yy-008
• Pressure sensor
(P1, bottom)
• Pressure sensor
(P2, middle)
• Pressure sensor
(P3, top)
P1
Tank Side
Monitor
P1 Position
P1
Tank Side
Monitor
P1 Position
P3
P3 Position
P1
Tank Side
Monitor
P1 Position
P2
P1-P2 Distance
P1
Tank Side
Monitor
P1 Position
P2
P1-P2 Distance
P3
P3 Position