Instruction manual

Commissioning
62 Installation and Operations Manual
New point
One can enter a new point. Suggested values displayed for the measured value and dip
value are the current uncorrected level or remaining fill height. The new value pair can
be altered without selecting the “edit point” parameter. If the table is full, you can still
select this parameter. In this case, the number of free table entries stands at minimum
value (= 0).
Edit point
The displayed value pair can be changed. Only the dip value can be changed with
semi-automatic input mode.
Caution! To accept the value pair in the table, confirm it with “store point”.
Store point
The displayed value pair is sorted in the table.
Note For sorting, the following criteria must be met:
Measured variables may not be equal but have different dip values.
A measured variable available in the table is recognized as equal when it is
closer than 1 mm to the sorting value.
After successful sorting, the setting remains at “edit point” and the number of
free table entries is decremented.
Caution! If the value cannot be sorted, the setting remains at the previous menu
option. No warning or error message is generated. However, the number of
remaining table entries is not decremented.
Delete point
The currently displayed point is deleted from the table. After deletion, the previous point
is displayed. If the table only consisted of one point before deletion, then the current
measured variable is displayed as a value pair.
Return
By selecting this point, you return to the function “dip table mode” (033).
Next point
This scrolls down in the table. If the table is empty, you can still select this option.
However, the displayed value does not change.
Previous point
This scrolls up in the table. If the table is empty, one can still select this option. However,
the displayed value does not change.
Caution! After entering one or more points into the dip table, make sure that the
dip table is activated in the “table on” dip table mode.