Instruction manual
Commissioning
60 Installation and Operations Manual
Manual
The “manual mode” of the dip table can be used to enter collected data after a series of
data pairs taken at different tank levels. The parameter “manual” in the “dip table” (033)
function can be used to enter data pairs, which have been recorded at different levels,
into the dip table.
The measured value and the dip value can be entered.
• uncorrected measured value:
This is the measured value supplied by the device, not corrected by the dip table.
The choice of measured value, level, or remaining fill height is dependent on the
device setting.
• Dip value:
This is the level or distance to flange respectively, given by the hand dip. This value
should be used to correct the measured value.
Note The bigger the distance between the different levels while taking hand dips,
the more accurate the linearization of the dip table will be.
Semi-automatic
The value pairs in the dip table can be read. Enter the dip value only. When there are new
value pairs, the current uncorrected level or distance is accepted as the measured value.
Table on
The dip table is switched on.
Table off
The dip table is switched off.
Clear table
The complete dip table is deleted. The table is switched off. The number of free table
entries is set to the maximum value (= 32).
View
The value pairs in the dip table can only be read. Select this menu option even if there
is no dip table available. In this case, the number of free table entries is at maximum
value (= 32).