Instruction manual

Commissioning
56 Installation and Operations Manual
Caution! During the learning period, fast filling/emptying, or turbulent surfaces
can result in switching off and on the phase evaluation. Subsequently, observed
measurement errors will disappear as soon as tank levels come back to areas
measured by 7500 series RTG previously with activated phase evaluation. If the
observed measurement errors are corrected by dip table entries, the
7500 series RTG will take care of these corrections and automatically adjust the
index correction table. Do not correct any settings in the basic calibration or the
extended calibration.
Note Immediately after installation, the 7500 series RTG measures with the
specified mm-accuracy. Until the level range has been completely covered by the
medium (setting up the correction table), the maximum permissible filling speed is
100 mm level change / min. After this, the fill speed has no limitation.
Function
pipe diam. corr.” (032)
For level measurement in stilling wells, radar systems require highly precise pipe inner
diameter data.
An mm-exact level measurement cannot be guaranteed for deviations
from the actual stilling well inner diameter of more than ± 0.1 mm to the value entered
in the function group “basic setup” (00).
The errors which occur as a result are linear and
can be corrected with a dip table containing at least two entries. The 7500 series RTG
also has an automatic pipe inner diameter correction. This adjusts the entered stilling
well inner diameter (input in the function group “basic setup” (00)) to the actual values.
However, this presupposes that the value entered in the function group “basic setup”
(00) matches the actual pipe inner diameter as accurately as possible. The user-defined
value entered in the function group “basic setup” (00) can be corrected with this value.
Switch on the “pipe diam. corr.” (032) function, after a level change of at least 5 m has
occurred since start-up. The pipe diameter, which the device determines automatically,
will then be transmitted to the “pipe diameter” (007) function.
Note If, and only if the “pipe diameter” (007) function has changed its value, it is
necessary to perform a “history reset” (009) and to delete the dip table after
activation of the “pipe diam. corr.” (032) function. Otherwise the level change of 5 m
has not yet been exceeded. The “pipe diam. corr.” (032) function must be
deactivated again and the procedure should be repeated at a later point of time.
E
+