Instruction manual

Commissioning
44 Installation and Operations Manual
7.3 Basic Setup
Figure 7-1: Basic Setup1
SD
F
L
D
BD
E
Distance
correct
Meas. value
correct
Distance
incorrect
Meas. value
incorrect
flange:
referencepoint of
measurement
E = empty calibr. (= zero)
settings in 005 settings in 006
D = distance (distance flange / product) L
display in 0A5 in 0A6
F = full calibr. (= span)
= level
display
SD = safety settings BD = blocking dist.
settings in 015 settings in 059

(description see )
basic setup (standard)
option
Commissioning
start mapping
Check distance
Range of mapping
adjust
“empty calibr.”
IOM117F
tank shape
flat ceiling
measuring cond.
unknown
process cond.
standard
empty calibr.
full calibr.
pipe diameter
(for bypass/stilling well)
dist. / meas.
value
According to
envelope curve
According to
hand dipping
history reset
dip table
first point corrects offset
further points correct linearity