6.5 HART Communication
Apart from local operation, parameterization of the measuring device and viewing of the
measured values by means of a HART protocol can also be accomplished. There are two
options available for operation:
Operation via the universal handheld operating unit, the Field Communicator 375,
475
Operation via the Personal Computer (PC) using the operating program
(e.g. FieldCare; see Figure 5-4 on page 29).
Operation via the 4590 Tank Side Monitor.
Note The 7500 series RTGs can also be operated locally using the keys. If
operation is prevented by the keys being locked locally, parameter entry via
communication is not possible either.
6.5.1 Protocol specific data
Table 6-11: Protocol Specific Data
6.5.2 Operation with Field Communicator 375, 475
All device functions can be adjusted via a menu operation with the Field Communicator
375, 475.
Note Further information on the handheld unit is given in the respective operating
manual included in the transport bag of the Field Communicator 375, 475.
6.5.3 Operation with Varec operating program
FieldCare is a Varec asset management tool based on FDT technology. With FieldCare,
you can configure all Varec devices as well as devices from other manufacturers that
support the FDT standard.
Manufacturer ID 000011 hex
Device Type Code 0010 hex
Transmitter specific revi-
sion
03 hex
HART specification 5.0
DD-Files Information and files can be found on:
www.varec.com
www.hartcomm.org
Load HART Min. 250
Device variables Primary value: level or volume
1
Features supported Burst mode
Additional Transmitter Status
1. according to configuration