6.4 Display and Acknowledging Error Messages
Errors that occur during commissioning or measuring are displayed immediately on the
local display. If two or more system or process errors occur, the error with the highest
priority is shown on the display.
The measuring system distinguishes between the following types of error:
• A (Alarm):
Device goes into a defined state (e.g. MIN, MAX, HOLD)
Indicated by a constant symbol.
(For a description of the codes, see »System Error Messages« on page 78.)
• W (Warning):
Device continue measuring, error message is displayed.
Indicated by a flashing symbol.
(For a description of the codes, see »System Error Messages« on page 78.)
• E (Alarm / Warning):
Configurable (e.g. loss of echo, level within the safety distance)
Indicated by a constant/flashing symbol.
(For a description of the codes, see »System Error Messages« on page 78.)
Table 6-10: Type of Errors
6.4.1 Error messages
Error messages appear as four lines of plain text on the display. In addition, a unique
error code is also output. A description of the error codes, see »System Error Messages«
on page 78.
• The “diagnostics” (0A) function group can display current error as well as the last
error that occurred.
• If several current errors occur, use
to page through the error messages.
• The last occurring error can be deleted in the “diagnostics” (0A) function group
with the function “clear last error” (0A2).
E
+
–
+
-
or