Use Instructions

9 of 20 UD00023.01
THIS DOCUMENT AND DATA ARE ISSUED IN STRICT CONFIDENCE AND ARE NOT TO BE
REPRODUCED IN WHOLE OR IN PART WITHOUT WRITTEN PERMISSION FROM MANAGEMENT
Form TMP_PRD0002.05 Released December 12, 2019
10. Repair
Manufacturer Repair
1) Repairs are only to be performed by authorized service personnel. Ultradent to provide service personnel
with documentation to perform repairs.
Warranty
Ultradent hereby warrants that this instrument shall, for a period of 5 years from the date of purchase*,
conform in all material respects to the specifications therefore as set forth in Ultradent’s documentation
accompanying the product and be free from any defects in materials/or workmanship. This warranty applies
solely to the original purchaser and is not transferable. All defective products are to be returned to Ultradent.
There are no user service components of the VALO curing light system. Tampering with the VALO curing light
will void its warranty.
The VALO curing light warranty does not cover customer damage. For example, if a VALO curing light is misused
or dropped and the lens breaks, the customer will be responsible to pay for any necessary repairs.
*With sales receipt indicating the date of sale to the dentist.
11. Processing
If the product is reusable or requires processing before use, use the following language (if applicable) or fill out
the below table to be compliant with ISO 17664. (UD Only)
To avoid cross-contamination, wipe syringe with an intermediate level disinfectant between uses and use a
disposable barrier sleeve. If a barrier sleeve is not used, syringe should be treated as single-use.
VALO Curing Light Processing Instructions
Disinfection Reprocessing Instructions
Initial treatment at the time of use
Can be disinfected in the event the device becomes
contaminated during use. A new disposable barrier sleeve
must be used for each patient.
Preparation before disinfection
Use a 70% isopropyl alcohol (IPA) wipe to thoroughly wipe
the entire surface and remove any debris. Discard soiled
wipe.
Rinsing
None
Drying
Allow to completely air dry at ambient temperature.
Maintenance, inspection, and testing
Visually inspect that the device is not damaged and is free of
debris. If debris is present, repeat preparation step with a
new wipe.
Packaging
No packaging required for disinfection.
Disinfection
Wipe the entire surface and keep wet for the exposure time
while paying special attention to cracks, crevices, seams, and
hard-to-reach areas. Use additional wipes as necessary to
keep the surface wet.
Disinfectant
Exposure Time
70% IPA wipe
4 minutes
Storage
Store in a clean and dry location. To prevent moisture
buildup, do not store in a barrier sleeve.