Use Instructions
8 of 20 UD00023.01
THIS DOCUMENT AND DATA ARE ISSUED IN STRICT CONFIDENCE AND ARE NOT TO BE
REPRODUCED IN WHOLE OR IN PART WITHOUT WRITTEN PERMISSION FROM MANAGEMENT
Form TMP_PRD0002.05 Released December 12, 2019
Charge State Description
Charge Status Indicator
Voltage Condition
Charge Current
Pre-Charge
Solid Amber
2.86 Vdc to 3.15 Vdc
95 mA
Bulk Charge
Solid Amber
3.15 Vdc to 4.20 Vdc
950 mA
Charge Complete
Solid Green
>4.20 Vdc
N/A
Standby
Both off
N/A
N/A
• Cord Adapter - Remove cap at the base of the curing light with a counter-clockwise turn. Remove
battery/cell. Insert cord adapter into battery/cell compartment, fully seat adapter and turn clockwise to lock
into place.
Warning Indicators for VALO X Curing Light
• Red Flashing = Low Battery – Should have about 10 curing cycles after indication.
• Red Steady = Dead Battery
• Amber indicates an over temperature state of the device.
• Flashing Red and Green indicate AF lock out.
Mounting Bracket Instructions
1) Bracket should be mounted to a flat, oil-free surface.
2) Clean surface with rubbing alcohol.
3) Peel backing off the bracket’s adhesive tape.
4) Position bracket so the curing light lifts upward when removed. Press firmly into place.
5) If running the curing light on battery/cell power, the curing light should be oriented downward in
the bracket. If running the curing light on the VALO X Cord Adapter power, the curing light should
be oriented upward in the bracket.
9. Maintenance
User-Performed Maintenance
1) End of life is determined by wear and damage due to use. Inspect all components for damage prior to use
and contact authorized service personnel to repair damage.
2) Routinely check the lens for cured dental resins. If necessary, use a plastic or stainless steel dental
instrument to carefully remove any adhered resin.
3) Routinely check the output in Standard Power mode. Light meters differ greatly and are designed for
specific light guide tips and lenses. NOTE: the true numeric output will be skewed due the inaccuracy of
common light meters and the custom LED pack in the curing light; however, contact authorized service
personnel if a decrease in output is observed when measuring with the same meter.