Use Instructions
4 of 20 UD00023.01
THIS DOCUMENT AND DATA ARE ISSUED IN STRICT CONFIDENCE AND ARE NOT TO BE
REPRODUCED IN WHOLE OR IN PART WITHOUT WRITTEN PERMISSION FROM MANAGEMENT
Form TMP_PRD0002.05 Released December 12, 2019
• Use caution when treating patients who suffer from adverse photobiological reactions or sensitivities,
patients who are undergoing chemotherapy treatment, or patients being treated with photosensitizing
medication.
• This unit may be susceptible to strong magnetic or static electric fields, which could disrupt the
programming. If you suspect this has occurred, unplug the unit momentarily and then re-plug it into the
outlet.
• DO NOT wipe down the curing light with caustic or abrasive cleaners, autoclave, or immerse in any kind of
ultrasonic bath, disinfectant, cleaning solution, or liquid. Failure to follow included processing instructions
may render curing light inoperable.
• To avoid damaging the equipment, DO NOT insert fingers, instruments, or other objects into the battery/cell
compartment of the curing light.
• To avoid damaging the equipment, DO NOT attempt to clean the gold contacts, or any part of the battery/cell
compartment. Call Ultradent Customer Service if there is a concern.
• To help prevent cross contamination and help keep dental composite material from adhering to the surface
of the lens and wand body, a barrier sleeve must be used over the VALO X curing light with each use. Barrier
sleeves are intended for single-patient use.
• To reduce the risk of corrosion, remove barrier sleeve after use.
• To reduce the risk of under-cured resins, do not use curing light if lens is damaged.
• To prevent the risk of product damage, do not autoclave or dry heat sterilize the VALO X Accessories.
• DO NOT use any VALO X Accessories for complete cure.
• DO NOT use illumination/visualization accessories for curing dental restorative materials and adhesives.
Country- /Region- Specific (UD Only)
Enter Country/Region (CA)
• WARNING: This product can expose you to substances which are known to the State of California to cause
cancer. For more information, go to www.P65Warnings.ca.gov
8. Stepwise Instructions
NOTE: Prior to each use, place a new barrier sleeve over the curing light, minimizing wrinkles over the lens for
best results.
Curing Modes
Curing Mode
Power
(mW)
Irradiance
(mW/cm2)
Total
Exposure
Time
(Seconds)
Energy
(Joules)
Button
Indicator
Standard Power
1,350
1,100
10
13.5
Steady Blue
Xtra Power - Triple Pulse
2,700
2,200
5
13.5
Pulsing Blue
NOTE: Xtra Power - Triple Pulse: 0.5 second auto rest between pulses with 2 second cooling period after
completion of the triple pulse.
NOTE: For optimal results, position curing light centered over and as close to resin as possible without
contacting surface.
Changing between modes:
Drum Tap (AF = accelerometer function) air forward to change through 2 curing modes: