6. Contraindications
Universal (UD Only)
• No known contraindications have been identified for this product.
Chemical (UD Only)
• For patients or users with allergy concerns, refer to product allergen document available at
www.ultradent.com. If allergic reaction is observed, rinse exposed area thoroughly with water and have the
patient consult their physician.
Country- /Region- Specific (UD Only)
Enter Country/Region (UD Only)
• Add contraindications here.
7. Warnings and Precautions
Universal (UD Only)
• CAUTION Possibly hazardous optical radiation emitted from this product. DO NOT look directly into the
curing light output. Patient, clinician, and assistants should always wear amber colored UV eye protection
when curing light is in use.
• To prevent the risk of electric shock, no modification of this equipment is allowed. Use only the included
Ultradent VALO power supply and plug adapters. If these components are damaged, do not use and call
Ultradent Customer Service to order a replacement.
• Portable RF communications equipment may degrade performance if used closer than 30 cm (12 in).
• Use only authorized accessories, cables, charger, batteries, and power supplies to prevent improper
operation, increased electromagnetic emissions or decreased electromagnetic immunity (refer to
Electromagnetic Emissions section).
• To avoid the risk of electrical fire associated with handling of batteries:
o DO NOT autoclave or spray battery, battery contacts, charger, or AC power adapter with
liquid of any kind. If corrosion appears on the contacts of the smart charger, call Ultradent
Customer Service to order a replacement.
o DO NOT charge batteries around flammable materials
o DO NOT keep charger in clinical operatory
o DO NOT charge non-rechargeable batteries/cells
o DO NOT recharge Ultradent batteries with a non-Ultradent charger
• To avoid the risk of injury, DO NOT use batteries that are corroded (rust), dented, emit an odor or fluids, have
a torn or missing wrapping, or are otherwise damaged. Call Ultradent Customer Service to order replacement
batteries/cells.
• To prevent the risk of thermal irritation or injury, avoid back-to-back curing cycles and do not expose oral soft
tissues at close proximity for more than 10 seconds in Standard Power mode or 5 seconds in Xtra Power -
Triple Pulse mode. If longer curing times are required, use multiple shorter curing cycles or use a dual-cure
product to avoid heating soft tissue.