Use Instructions

16 of 20 UD00023.01
THIS DOCUMENT AND DATA ARE ISSUED IN STRICT CONFIDENCE AND ARE NOT TO BE
REPRODUCED IN WHOLE OR IN PART WITHOUT WRITTEN PERMISSION FROM MANAGEMENT
Form TMP_PRD0002.05 Released December 12, 2019
IEC 61000-4-4
± 1 kV for
input/output lines
Note 1: The curing
light has no I/0
ports
The curing light is battery/cell
powered and is not capable of
connection to AC MAINS power.
Because the curing light is
battery/cell powered, it is not subject
to electric transients, surges, voltage
dips, shorts, interruptions, or
variations on AC MAINS power.
The accessory battery charger may
be subject to the above but it is
separate from and not critical to the
operation of the VALO X curing light.
If battery voltage of the curing light
drops to 4VDC the unit will not allow
operation. The VALO X curing light
will turn off. When new batteries are
put in and the proper power levels
are restored, the curing light will
restart and return to the same state
before power loss. The curing light
will self-recover in the event of
power loss.
Surge
IEC 61000-4-5
± 1 kV line to line
± 2 kV line to earth
± 1 kV line to line
± 2 kV line to earth
Voltage, dips,
shorts,
interruptions
and variations
on the power
supply input
lines
IEC 61000-4-11
<5% U
(>95% dip in U for
0.5 cycle)
40% U
(60% dip in U for 5
cycles)
70% U
(30% dip in U for 25
cycles)
<5% U
(>95% dip in U for 5
s)
<5% U
(>95% dip in U for
0.5 cycle)
40% U
(60% dip in U for 5
cycles)
70% U
(30% dip in U for
25 cycles)
<5% U
(>95% dip in U for
5 s)
Note 2: Self
recovers
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical,
residential, home health care,
commercial, hospital, or military
environment.
Note 1: The curing light is not equipped with any ports or any accessible I/O lines.
Guidance and Manufacture’s Declaration for Electromagnetic Immunity for non-life support systems
The curing light is intended for use in the electromagnetic environment specified below. The customer or
user should ensure that it is used in such an environment.
IMMUNITY
test
IEC 60601 test
level
Compliance
level
Electromagnetic environment guidance
Conduction RF 3 Vrms 3 Vrms
Portable and mobile RF communications
equipment should be used no closer to any
part of the curing light, including cables, than
the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.