Use Instructions
15 of 20 UD00023.01
THIS DOCUMENT AND DATA ARE ISSUED IN STRICT CONFIDENCE AND ARE NOT TO BE
REPRODUCED IN WHOLE OR IN PART WITHOUT WRITTEN PERMISSION FROM MANAGEMENT
Form TMP_PRD0002.05 Released December 12, 2019
14. Miscellaneous Information
Guidance and Manufacture’s Declaration for Electromagnetic Emissions
The curing light is intended for use in the electromagnetic environment specified below. The customer or user
should ensure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The curing light uses a cell and is not affected by EMI, RF, or
surge suppression.
RF emission
CISPR 11
Class B
The curing light uses electrical and electromagnetic energy only
for their internal functions. However, any RF emissions are very
low and are not likely to cause interference in nearby electronic
equipment.
Harmonic
emissions
IEC 61000-3-2
N/A
Harmonic emission and voltage fluctuation testing are not
applicable to the curing light because it is cell/battery powered.
Voltage
fluctuations/flicker
emissions IEC
61000-3-3
N/A
The curing light is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings for domestic use.
Guidance and Manufacture’s Declaration for Electromagnetic Immunity
The curing light is intended for use in the electromagnetic environment specified below. The customer or user
should ensure that it is used in such an environment.
IMMUNITY test
IEC 60601 test level
Compliance level
Electromagnetic environment
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Physical environment should be
restricted to the following:
1. IP Code: IP20
2. Do not immerse in liquid.
3. Do not use around flammable
gas. Unit is Non-APG and Non-AP.
4. Storage humidity range: 10% -
95%
5. Storage temperatures range:
10° C - 40° C
Electrical fast
transient/burst
± 2 kV for power
supply lines
± 2 kV for power
supply lines