NOTE: PointCure and ProxiCure lenses focus the light energy into a smaller aperture, thus increasing
the light intensity above calibrated values.
VALO X
Curing Light
Ratings: IEC 60601-1 (Safety), IEC
60601-1-2 (EMC)
Weight:
With battery: 4.6 oz. (137 grams)
Without battery: 3.6 oz. (103 grams)
With cord adapter: 5.6 oz. (158 grams)
Dimension (battery configuration): 8.64 x 0.83 x 0.83
inches, (219 x 21 x 21 millimeters)
Power
Supply
Output - 9VDC at 2A
Input - 100VAC to 240VAC, 50-60 Hz
Ultradent P/N 4952 - VALO X Power
Supply with international plug inserts
Ratings: IEC 60601-1 (Safety)
Cord Length - 6 feet (1.8 meters)
VALO Power Supply is a Medical Grade Class II power
supply and provides isolation from MAINS power
VALO Smart
Charger
VALO X 3.7VDC Lithium Ion (Li-Ion) smart charger:
Automatic shut off when fully charged
Auto-detection of defective cell
Protections: Overcharge, Short-circuit, Reverse polarity
Yellow LED - Charging
Green LED - Empty or Fully Charged
Charging time: 1–3 hours
Rating: CE, WEEE
VALO cells
Rechargeable: Li-ion 15/740
Working Voltage: 3.7VDC
Ratings: WEEE, IEC 62133-2
Operating
Conditions
Temperature: +10°C to +32°C (+50°F to +90°F)
Relative Humidity: 10% to 95%
Ambient Pressure: 700 hPa to 1060 hPa
Duty Cycle:
At maximum ambient temperature (32°C) 10 seconds ON and 60 seconds OFF.
Trouble Shooting
If the solutions suggested below do not rectify the problem, please call Ultradent at 800.552.5512. Outside
the United States, call your Ultradent distributor or dental dealer.
Problem
Possible Solutions
Light will not turn on
Press the Power Button to wake from Power Save Mode
Confirm battery/cell is fully charged
Check that cell is correctly inserted into the unit
If amber Warning LED is flashing this means the curing light
has reached its internal temperature safety limit. Allow the
curing light to cool down for 10 minutes or use a cool
moist towel to cool the unit down quickly.