Problem 
Probable Cause 
Remedy 
Inconsistent serial 
communication or missing 
data. 
Pin contacts of the connector have 
become oxidized or the connector is 
not firmly attached. 
Change the WMT700 cable. 
Retrofit installation of 
WMT700 is not possible. 
You do not have the correct mounting 
kit for the retrofit installation. 
Check that you are using the 
correct mounting adapter with 
your mounting kit. Refer to 
Appendix F, Accessories, on 
page 213. 
WMT700 does not work 
properly after retrofit 
installation. 
Incorrect cables were used in the 
retrofit installation. 
Check that you have used cables 
listed in Checklist for Connection 
Cables  on page 73. 
If you do not have the correct 
cables, contact Vaisala technical 
support. 
Connection to WMT700 is 
lost. 
The power supply is not sufficient, 
especially if you are using a heated 
version of WMT700. 
Check that the power supply 
matches the requirements in 
section Powering on page 105. 
Error and Event Messages 
If messages parameter is set to 1, WMT700 sends error and event 
messages. Table 56 below lists the available messages. 
Table 56  Error and Event Messages 
Error/Event Code 
Probable Cause 
Remedy 
2 
Event. 
Parameters have been set to factory 
defaults. 
You can change the default 
settings in configuration mode; 
see 
Communicating with 
Terminal Software on page 107.  
3 
Event. 
Wind calibration data has been lost. 
WMT700 needs to be calibrated. 
Contact Vaisala technical 
support. 
10 
Error. 
You have given an invalid value when 
using the S command. 
Check the allowed parameter 
values; see Appendix D 
Configuration Parameters 
on 
page 205. 
11 
Error. 
Parameter used with the S command 
is unknown. 
Check that you are using the 
correct parameter names, see 
Appendix D Configuration 
Parameters on page 205. 
12 
Error. 
Unknown command.  
Check the allowed commands in 
Appendix A, Complete Command 
Set for WMT700, on page 195.  
13 
Error. 
String before the end of command 
character is too long. 
Check the command and 
parameter names that you are 
using.  