Periodic Maintenance 
WMT700 is a very reliable and rugged sensor. Since no moving 
or consumable parts are used, no periodic maintenance is 
required. WMT700 has been calibrated at the factory, and you are 
not required to perform recalibration. 
NOTE 
Periodic 
calibration of WMT700 is not required for technical reasons. 
You can use the WM Verifier to test the mechanical integrity of the 
device and perform one
-point calibration. 
However, some quality management systems may require regular 
calibration of the measu
ring instruments. 
To fulfill these requirements, Vaisala recommends that you recalibrate 
the wind sensor every 24
 months. Contact Vaisala Service Center for 
more information.
Visual Inspection 
Vaisala recommends that you carry out a periodic visual 
inspection of the wind sensor to ensure correct operation and 
clean the wind sensor if it becomes contaminated. If necessary, 
you can also test the operation of WMT700 with an optional 
verifier, which is a small echo-free chamber. 
If any of the transducers have been bent, twisted or rotated, the 
measurement results might not be accurate. 
To avoid errors caused by a damaged array, check the following: 
-  Array has not been hit or bent. All transducers must be parallel 
to each other. 
-  Transducers have not been scraped or touched with sharp 
objects. The silicon rubber transducer sleeve must be 
undamaged. 
Cleaning 
Regular cleaning of WMT700 is not required. If the wind sensor 
becomes contaminated, you can clean it with a soft cloth 
moistened with a mild detergent. Do not use solvents or a 
pressure washer to clean WMT700, since they may damage the 
silicon rubber transducer sleeve. 