User`s guide

Chapter 5 _________________________________________________________________ Operation
To start continuous measurement, use the START command.
You can stop the measurement with the STOP command. For
instructions, refer to START Start Continuous Measurement
and STOP Stop Wind Measurement on page 115.
WS425 A/B NMEA Extended
Commands
This command polls data from WMT700.
$WIP<id>Q,*<chk><CR><LF>
where
$WIP
=
Fixed text
<id>
=
Data ID; A … Z
Q
=
Fixed text
*
=
Fixed text
<chk>
=
Checksum (8-bit XOR, excluding $ and *)
<CR>
=
Carriage return code, ASCII 0DH
<LF>
=
Line feed code, ASCII 0AH
Table 48 Checksum Table
ID Character <id>
Checksum <chk>
Polling String
A
72
$WIPAQ,*72<CR><LF>
B
71
$WIPBQ,*71<CR><LF>
C
70
$WIPCQ,*70<CR><LF>
D
77
$WIPDQ,*77<CR><LF>
E
76
$WIPEQ,*76<CR><LF>
F
75
$WIPFQ,*75<CR><LF>
G
74
$WIPGQ,*74<CR><LF>
H
7B
$WIPHQ,*7B<CR><LF>
I
7A
$WIPIQ,*7A<CR><LF>
J
79
$WIPJQ,*79<CR><LF>
K
78
$WIPKQ,*78<CR><LF>
L
7F
$WIPLQ,*7F<CR><LF>
M
7E
$WIPMQ,*7E<CR><LF>
N
7D
$WIPNQ,*7D<CR><LF>
O
7C
$WIPOQ,*7C<CR><LF>
NOTE
To use WMT700 with the NMEA Extended profile, either set the
autoSend
parameter to 0 to enable polling or define a fixed output
interval with the
autoInt parameter. If you are using automatic messages,
the value for the
autoSend parameter must be set to 15.
VAISALA _______________________________________________________________________ 157