User`s guide
Chapter 5 _________________________________________________________________ Operation 
Entering and Exiting Configuration Mode 
With OPEN and CLOSE commands you can toggle between 
configuration mode and measurement mode. In the measurement 
mode, OPEN changes the mode to the configuration mode, and in 
the configuration mode, CLOSE changes the mode to the 
measurement mode. Note that the configuration mode does not 
recognize the OPEN command and the measurement mode does 
not recognize the CLOSE command. 
OPEN — Entering Configuration 
Mode 
To switch WMT700 from the measurement mode to the 
configuration mode, use the CLOSE command. 
$0OPEN<enter> 
|| |  |____
To activate the command, press Enter 
|| |_________The OPEN command 
||____________
Sensor address, all sensors answer 0 address 
|_____________Fixed $ character starts command 
After a successful OPEN command the sensor switches from 
measurement mode to configuration mode and the following 
symbol is displayed: 
> 
CLOSE — Exiting Configuration 
Mode 
To switch WMT700 from the configuration mode to the 
measurement mode, use the CLOSE command. 
>CLOSE<enter> 
| |  |____
To activate the command, press Enter 
| |__________
The CLOSE command 
|_____________
Configuration mode prompt 
VAISALA _______________________________________________________________________ 109 