At the end of each command, press ENTER to execute the command. For 
a successful execution, the following combinations of Carriage Return 
<CR> and Line Feed <LF> are accepted: 
- <CR> 
- <LF> 
- <CR><LF> 
You can send commands at any Baud rate. The command is executed 
when end-of-line character is received. 
If messages have been enabled, WMT700 responds to an invalid 
command with an error message. If the messages have been disabled, 
WMT700 does not send any error messages. You can fetch the most 
recent error data in configuration mode with the ERRORS command, 
see section ERRORS — Get Error Codes and Counts on page 78. 
NOTE 
You can configure WMT700 by sending individual commands or by 
loading a configuration file to the sensor. For instructions on configuring 
all settings at once using configuration files, see section User-
Configurable Data Messages on page 83 
Table 17  List of Configuration Mode Commands 
Command  Description 
?  Displays a list of configuration commands. 
BAUD  Changes or displays serial port settings. 
CLEARERR Resets error counters. 
CLOSE  Switches the serial port to measurement mode. 
ERRORS  Fetches the error codes and counter information from WMT700.
G  Displays either all or specified parameters. 
H  Displays list of data messages and available values for 
measurement unit, profile, baud rate, interface, and analog 
output mode. 
MEAS  Starts wind measurement based on the user-configurable 
averaging time. WMT700 does not send data messages 
automatically. 
POLL  Tests data polling. 
RESET Resets WMT700. 
S  Changes selected parameters or defines new data messages. 
START  Starts continuous measurement. 
STOP  Stops continuous measurement. 
VERSION  Displays the software version. 
WIND_GET  Fetches wind calibration information. 