User`s guide

Chapter 9 ___________________________________________________________ Troubleshooting
VAISALA ______________________________________________________________________ 113
11. Updating the settings will be completed in a few seconds. When the
operation is done, WMT700 sends the following response:
Done. Rebooting…
12. WMT700 has now applied the settings and will reboot using the
new settings. The restored settings are shown in Table 27 below.
13. Before you start configuring new values to WMT700 you need to
apply the new settings to HyperTerminal. From the Call menu,
click Disconnect to close the connection.
14. From the File menu, click Properties.
15. In the Properties dialog, click Configure.
16. Select 9600 in the Bits per second combo box.
17. Close the Properties window.
18. From the Call menu, click Call.
Table 27 Restored Serial Port Settings
Parameter
Name
Default Value Description
com1_baud
com2_baud
4 4=9600 bauds per second
com1_data
com2_data
8 Number of data bits
com1_delay
com2_delay
20 Response delay in milliseconds
com2_interf 0 0=RS-485
com1_parity
com2_parity
0 0=None
com1_protocol
com2_protocol
0 WMT700 protocol
com1_stop
com2_stop
1 Number of stop bits
The serial port settings have now been restored to known values. You can
send the OPEN command and start configuring WMT700. For
information on the configuration commands, see section Configuration
commands on page 71.
NOTE
You can also use other terminal programs such as Tera Term.