User guide
Swapping Devices
viewLinc 4.0 Administrator Guide 43
Note: Each Vaisala Veriteq device must be reviewed
separately using vLog, and the COM port to
which they are connected must be selected.
• COM port/IP address. Device is on the same COM port
(attached to the same cable). If you are swapping a 300
Series Transmitter connected as a network device,
ensure the new transmitter has the same IP address.
•
Name. Device is not retired or deactivated.
•
Status. Device is not in active alarm mode.
•
Sample Interval. Devices have the same sample
interval.
•
Channels. Devices have the same channels enabled
and be using the same units of measure for each
matching channel.
•
Audit Trail. Vaisala Veriteq devices are not linked to a
vLog Audit Trail.
2 From Options | Locations, navigate the system tree to
find the device you want to swap. Right click and select
Pause Device Alarming.
3 In the Pause Device Alarming screen, you can add a
comment about why you are swapping devices, and
specify the expected duration for the pause. This will
ensure no communication or threshold alarms are
triggered when disconnecting the device.
4 Click OK.
5 Disconnect the device you want to swap, and connect
the new device. If you are swapping a 300 Series
Transmitter, unplug the power source and plug it in
again.