Manual

User's Guide _______________________________________________________________________
98 ___________________________________________________________________ M210764EN-F
Using Display/Keypad
Use display/keypad to select the display output quantities.
1. Press any of the arrow buttons to open the Main Menu.
2. Press the arrow button to select Display.
3. Press arrow button to select Quantities.
4. Select the quantity by using the ▲▼ arrow buttons. Confirm the
selection by pressing SELECT. You can select 1 ... 3 display
quantities at a time.
5. Press EXIT to return to the basic display.
To select display units:
1. Press any of the arrow buttons to open the Main Menu.
2. Press the arrow button to select Display.
3. Use the ▲▼arrow buttons to select Units. Confirm the selection
by pressing the right-hand arrow button.
4. Use the ▲▼ arrow buttons to select display units. Confirm the
selection by pressing CHANGE. The unit changes from metric to
non-metric or the other way round.
5. Press EXIT to return to the basic display.
NOTE
Changing the display quantities/units (by using the display/keypad) has
no effect on the serial output data.
Using Serial Line
FORM
Use the serial line command FORM to change the format or select a
certain quantities for the output commands SEND and R.
FORM [x]<cr>
where
x
=
Formatter string
If no formatter string is entered, the command will display the currently
active formatter string. Note that the hash symbol "#" is shown as a
backslash "\" when viewing the current formatter string.
Formatter string consists of quantities and modifiers. The MMT330
measures the following quantities: