humidity / dew point / TEMPERATURE / moisture in oil / carbon dioxide / pressure / wEATHER
Measure and Succeed Accurate measurements for humidity, temperature, dew point, carbon dioxide, pressure and moisture in oil If the success of your operations depends on faultless measurements, we have the solution for you: Vaisala HUMICAP®, DRYCAP®, CARBOCAP® and BAROCAP® sensors – the unrivaled industry quality benchmarks. Only true knowledge matters and only the world’s most accurate instruments can deliver it reliably. Based on over 75 years of experience.
Demanding measurement requirements across multiple industries, from semiconductor to automotive to power, have always inspired Vaisala to continuously develop measurement instruments that set the benchmark for quality, accuracy, and reliability. To ensure we continue to understand your needs, we keep a close eye on our customer surveys and on comments collected from the field.
Table of Contents by Parameters Humidity Vaisala HUMICAP® Sensor for Measuring Relative Humidity. . .................................................................................................. 4 How to Choose the Right Instrument for Measuring Humidity and Dew Point.......................................................................... 6 HMT330 Series Humidity and Temperature Transmitters for Demanding Humidity Measurement...........................................
Humidity Moisture in Oil Vaisala HUMICAP® Sensor for Measuring Moisture in Oil..................................................................................................... 110 MMT330 Moisture and Temperature Transmitter Series for Oil............................................................................................ 112 MMT310 Series Moisture and Temperature Transmitter for Oil. . ...........................................................................................
Humidity / TECHNOLOGY DESCRIPTION Vaisala HUMICAP® Sensor for Measuring Relative Humidity In 1973, Vaisala introduced HUMICAP ®, the world’s first thinfilm capacitive humidity sensor. Since then, Vaisala has become the market leader in relative humidity measurements, and thin-film capacitive humidity sensors have developed from one company’s innovation into a global industry standard.
Vaisala INTERCAP® Sensor ▪ Same measurement principle as in the HUMICAP® sensor ▪ Factory pre-calibrated - no additional calibration or adjustment needed ▪ ▪ Fully interchangeable Humidity In building automation, optimizing both the temperature and relative humidity of the indoor environment enables a greater level of comfort for occupants than optimizing temperature alone.
How to Choose the Right Instrument for Measuring Humidity and Dew Point What is Humidity? An Introduction to Humidity Parameters Partial Pressure of Water Vapor Humidity is simply water in its gaseous phase, properly called water vapor. Because water vapor is a gas, most of the common gas laws apply to it, including Dalton's law of partial pressures.
Effect of Pressure on Humidity Dalton's law states that a change in the total pressure of a gas must have an effect on the partial pressures of all of the component gases, including water vapor. If, for example, the total pressure is doubled, the partial pressures of all component gases are doubled as well. In air compressors, a pressure increase "squeezes" water out of the air as it is compressed.
Mixing Ratio The mixing ratio (x) is the ratio of water vapor mass to the mass of dry gas. It is dimensionless but often expressed in grams per kilogram of dry air. The mixing ratio is mainly used in drying processes and HVAC applications for calculating water content when the mass flow of air is known. Wet Bulb Temperature Traditionally, the wet bulb temperature (Tw) is the temperature indicated by a thermometer wrapped in a wet cotton sheath.
The Right Instrument for High Humidity In many applications it is advisable to isolate the probe from the process with a ball valve to enable the removal of the probe for maintenance without shutting down the process (see figure 3). Environments with >90%RH are defined here as high humidity environments. At 90%RH a difference of 2°C can cause water to condense on the sensor, which in an unventilated space may take hours to dry. Vaisala humidity sensors will recover from condensation.
In a pressurized system a sample pump isn’t needed as the process pressure induces a large enough flow to the sampling cell. When measuring dew point with a sampling system, trace heating should be used when the ambient temperature around the cooling coil or connecting tube is within 10°C of the dew point temperature. This prevents condensation in the tubing that connects the dew point instrument to the process.
Humidity manufacture and the measurement on site – introduces measurement error. The sum of these potential errors is the uncertainty of the measurement. When selecting a humidity sensor, consider the following factors associated with accuracy: • Linearity over the working range • Hysteresis and repeatability • Stability over a period of time • Temperature dependence of the sensor Humidity sensors from Vaisala.
cleaning agents. The chemical purge heats the sensor at regular intervals to remove chemicals that may have accumulated over time. The Importance of Electromagnetic Compatibility (EMC) There are many standards defining the ability of products to withstand external electrical interference. In addition, the product must not generate emissions that can cause interference to sensitive equipment.
Humidity Humidity Instrument Calibration In calibration, the humidity reading of an instrument is compared against a portable reference. The reference should be regularly calibrated and provided with a valid certificate. When selecting one of the many calibration methods, time, cost, technical requirements, expertise, and the unique needs of the organization must be balanced.
www.vaisala.com HMT330 Series Humidity and Temperature Transmitters for Demanding Humidity Measurement Features/Benefits The HMT330 transmitter family offers reliable performance for a wide variety of demanding industrial humidity measurements. The Vaisala HUMICAP® Humidity and Temperature Transmitter Series HMT330 is designed for demanding industrial applications where stable measurements and extensive customization are essential.
The wide variety of measurement probes, several installation accessories, and universal mains and DC power options make the instruments easy to install in various locations and kinds of environment; walls, poles, pipelines, and ducts, for example. The input/output cable can be fed through the back of the transmitter, which is a useful feature, especially for cleanroom installations.
Technical Data Performance Inputs and Outputs Relative humidity Operating voltage 10 ... 35 VDC, 24 VAC ±20% with optional power supply module 100 ... 240 VAC, 50/60 HZ Power consumption at +20 °C (Uin 24 VDC) RS232 max. 25 mA max. 25 mA Uout 2 x 0 ... 1 V/0 ... 5 V/0 ... 10 V max. 60 mA Iout 2 x 0 ... 20 mA display and backlight + 20 mA during chemical purge max. 110 mA during probe heating (HMT337) + 120 mA Analog outputs (2 standard, 3rd optional) current output 0 ... 20 mA, 4 ...
Humidity 77 (3.03) 27.5 (1.08) 71 (2.80) Dimensions 53.5 (2.10) 38 (1.50) 38 (1.50) Dimensions in mm (inches) 53.5 (2.10) 183 (7.20) 169 (6.65) ø 7 (0.28) 96 (3.78) 116 (4.57) M20 x 1.5 for cable diameter 8 ... 11 mm/0.31 ... 0.43" Conduit fitting 1/2" NPT User cable connector (optional) M12 series 8-pin (male) option 1 female plug with 5 m (16.4 ft.) black cable option 2 female plug with screw terminals Probe cable diameter HMT333 (+80 °C) 6.0 mm other probes 5.
HMT331 Humidity and Temperature Transmitter for Demanding Wall-Mounted Applications Typical Applications The HMT331 is a state-of-the-art wall-mounted humidity measurement instrument. The Vaisala HUMICAP® Humidity and Temperature Transmitter HMT331 is a high-quality wall-mounted transmitter for demanding HVAC and condition-monitoring applications.
Humidity HMT334 Humidity and Temperature Transmitter for High Pressure and Vacuum Applications Typical Applications ▪ ▪ test chambers high-pressure and vacuum processes Technical Data Temperature measurement range Operating pressure -70 ... +180 °C (-94 ... +356 °F) 0 ... 10 MPa (0 ... 100 bar) Accessories The HMT334 is ideal for permanent installations in pressurized or vacuum processes.
HMT337 Humidity and Temperature Transmitter for High Humidity Applications Typical Applications ▪ professional meteorology ▪ intake air monitoring of engines and gas turbines ▪ timber drying kilns Duct installation kit for HMT333 and HMT337. The HMT337 is ideal for the most demanding process and meteorological measurements in high-humidity condensing environments.
Humidity HMT338 Humidity and Temperature Transmitter for Pressurized Pipelines Typical Applications ▪ ▪ ▪ ▪ process lines environmental chambers vacuum-drying processes compressed air lines with refrigerant dryers The HMT338 is ideal for installations in pressurized processes where the probe needs to be removed while the process is running. The Vaisala HUMICAP® Humidity and Temperature Transmitter HMT338 is designed for pressurized processes.
Dimensions of the Probes for the HMT330 Series Dimensions in mm (inches) HMT335 probe 41 (1.6) HMT338 probe ø13.5 (0.53) HMT331 probe 51 (2.0) 192 (7.56) 253 (9.96) 37.5 (1.48) 78.5 (3.09) M9×1 ø 12 (0.47) HMT337 RH probe M9x1 Ø 12 (0.47) HMT333 probe 37.5 (1.48) 79.5 (3.13) 98.5 (3.88) 99.5 (3.92) Installation flange 50 (1.97) 5) .9 5 2 (0.08) ø7 ø12 (0.47) M26x1.5 (2 HMT337 T probe M22×1.5 or NPT 1/2" HMT334 probe 41 (1.61) 120 (4.72) 170 (6.69) ø9 7) 0.3 ( .
Humidity Looking for Accurate and Reliable Humidity Instruments? Only Vaisala HUMICAP® Measures Up. The quality and profitability of your process equals the quality of your measurement. The success of critical operations depends on small things, such as our humidity sensor. The Vaisala HUMICAP® technology has led the market for almost 40 years through unrivaled reliability continually introducing innovative features for better stability and usability.
www.vaisala.com HMT360 Series Intrinsically Safe Humidity and Temperature Transmitters Customers may specify the transmitter configuration when ordering the instrument, however changes in configuration can also easily be made in the field.
Dimensions Technical Data Dimensions in mm (inches) HMP361 for wall mounting Probe diameter 62 (2.44) -40 ... +60 °C (-40 ... +140 °F) ø12 (o.47) Temperature range Humidity Interchangeable Probes for HMT360 Intrinsically Safe Humidity and Temperature Transmitter 12 mm 41 (1.61) 86 (3.39) 127 (5) The HMP361 probe in this picture has a stainless steel netting filter. Dimensions Technical Data HMP363 for confined spaces Dimensions in mm (inches) -40 ... +120 °C (-40 ...
Dimensions HMP365 for high temperature Temperature range Probe cable length -70 … +180 °C (-94 … +356 °F) Dimensions in mm (inches) 41 (1.6) 2, 5 or 10 meters Probe diameter ø13.5 (0.53) Technical Data 51 (2.0) 192 (7.56) 13.5 mm 253 (9.96) Installation The HMP365 probe is designed for high temperature environments. 75 (2.96) 50 (1.97) Mounting flange 210696 Cable Gland M20x1.5 with splitting seal Installation flange: stainless steel HMP247CG drilling 16...22 (0.63...0.
Performance Operating Environment Relative humidity Temperature range operating temp. range for electronics with display storage Pressure range Measurement range 0 ... 100 % RH Accuracy (including non-linearity, hysteresis, and repeatability) for typical applications with Vaisala HUMICAP® 180R at +15 ... +25 °C (59 ... +77 °F) ± 1.0 % RH (0 ... 90 %RH) ±1.7 %RH (90 ... 100 %RH) at -20 ... +40 °C (-4 ... +104 °F) ±(1.0 + 0.008 x reading) %RH at -40 ... +180 °C (-40 ... +356 °F) ± (1.5 +0.
Classification with Current Outputs Dimensions Europe / VTT Dimensions in mm (inches) Safety factors II 1 G Ex ia IIC T4 Ga VTT 09 ATEX 028 X issue No: 2 Ui = 28 V, Ii = 100 mA, Pi = 700 mW Ci = 1 nF, Li negligibly low Environmental specifications -40 ... +60 °C (-40 ... +140 °F) Tamb 0.8 ... 1.
HMT361 Accessory Ball valve ISO 1/2 with welding joint Cable Gland M20 x 1.5 with splitting seal Duct installation kit Fitting body ISO1/2 solid structure Fitting body M22 x 1.
www.vaisala.com HMT310 Humidity and Temperature Transmitter The Vaisala HUMICAP® Humidity and Temperature Transmitter HMT310 models (from left to right): HMT313, HMT317, HMT314, HMT318, HMT315 and HMT311. Features/Benefits ▪ Latest generation Vaisala HUMICAP® sensor for excellent accuracy and stability ▪ Full 0 ...
Measured Values Electrical Connections RELATIVE HUMIDITY Measurement range Sensor Vaisala HUMICAP®180R Vaisala HUMICAP®180RC 0 ... 100 %RH typical applications applications with chemical purge/warmed probe catalytic sensor for H2O2 Vaisala HUMICAP®180VHP environments catalytic sensor with chemical Vaisala HUMICAP®180VHPC purge for H2O2 environments Accuracy (including non-linearity, hysteresis, and repeatability) at a temperature range of +15 ... +25 °C (+59 … +77 °F) ±1 %RH (0 ... 90 %RH) ±1.
43(1.69) 5 (0.20) Dimensions Transmitter body dimensions in mm (inches) Ø4.5 (0.177) 72.5 (2.85) 60.5 (2.38) 43(1.69) 5 (0.20) Ø4.5 (0.177) M22×1.5 or NPT 1/2" ø12 (0.47) M9x1 Ø 12 (0.47) 102 (4.02) 115 (4.53) 72.5 (2.85) 60.5 (2.38) Probe dimensions in mm (inches) 102 (4.02) 115 (4.53) 41 (1.61) 37.5 (1.48) 78.5 (3.09) 120 (4.72) HMT313 probe 98.5 (3.88) 170 (6.69) 51 (2.0) M9×1 ø 12 (0.47) 41 (1.6) ø13.5 (0.53) HMT314 probe 37.5 (1.48) 192 (7.56) 79.5 (3.13) 253 (9.96) 99.
Humidity Calculate and Convert. The free web based Vaisala Humidity Calculator allows you to calculate several humidity parameters from one known humidity value. Make unit conversions on the fly, and see the effects of changing ambient conditions, like temperature and pressure. The calculator is available in several languages. Access or download the Vaisala Humidity Calculator at www.vaisala.com/humiditycalculator Available also as: iPhone app Android app www.vaisala.
www.vaisala.com Vaisala HUMICAP® Humidity and Temperature Transmitters HMT120 and HMT130 Features/Benefits The HMT120/130 with and without a display. The Vaisala HUMICAP® Humidity and Temperature Transmitters HMT120 and HMT130 are designed for humidity and temperature monitoring in cleanrooms and are also suitable for demanding HVAC and light industrial applications. Performance The HMT120/130 incorporates Vaisala HUMICAP® technology that measures relative humidity accurately and reliably.
Performance Operating Environment RELATIVE HUMIDITY Operating temperature range transmitter body, no display transmitter body, with display HMP110 probe Storage temperature range Electromagnetic compatibility Measurement range 0 ... 100 %RH Accuracy including non-linearity, hysteresis, and repeatability temperature range 0 °C ... +40 °C 0 ... 90 %RH ±1.5 %RH 90 ... 100 %RH ±2.5 %RH temperature range -40 ... 0 °C, +40 ... +80 °C 0 ... 90 %RH ±3.0 %RH 90 ... 100 %RH ±4.
Dimensions Dimension in mm Remote probe dimensions 116 120 Wall Assembly dimensions 78 M4 ;4 pc s 71 [2,8] 91 145 39 46 ø12 [0,47] 72 92 Please contact us at www.vaisala.com/requestinfo www.vaisala.com 36 Ref. B211086EN-E ©Vaisala 2013 Scan the code for more information This material is subject to copyright protection, with all copyrights retained by Vaisala and its individual partners. All rights reserved. Any logos and/or product names are trademarks of Vaisala or its individual partners.
37 Humidity
www.vaisala.com HMW90 Series Humidity and Temperature Transmitters for High Performance HVAC Applications Features/Benefits The HMW90 Series Humidity and Temperature Transmitters are designed for demanding HVAC applications. Wall-mounted Vaisala HMW90 Series HUMICAP® Humidity and Temperature Transmitters measure relative humidity and temperature in indoor HVAC applications, where high accuracy, stability, and reliable operation are required.
Technical Data Models 2-wire, current output 3-wire, voltage output configurable analog output model 2-wire, current output 2-wire, current output with display 3-wire, voltage output 3-wire, voltage output with display Configurable analog/digital model Digital (BACnet, Modbus) model Digital (BACnet, Modbus) model with display Performance Relative humidity 0 ... 100 %RH, non-condensing +10 ... +40 °C (+50 ... +104 °F) ±1.7 %RH ±2.5 %RH -5 ... +10 °C, +40 ... + 55 °C (+23 ... +50 °F, +104 ... +131°F) 0 ...
www.vaisala.com HMD60/70 Humidity and Temperature Transmitters for Ducts in HVAC Applications Features/Benefits Vaisala HUMICAP Humidity and Temperature Transmitters HMD60 and HMD70 are designed for use in air conditioning applications where accurate and stable control of relative humidity and temperature are required.
60 Series: 2-wire, 4 to 20 mA Output Duct mount General - 60 Series RH only RH & T T only HMD60U HMD60Y HMD60T Supply voltage 10 ... 35 VDC (RL = 0 ohm) 20 ... 35 VDC (RL = 500 ohm) 4 ... 20 mA Output signal 70 Series: 3-wire, Variable Voltage Output General - 70 Series Duct mount Supply voltage range depends on the selected output signal DC AC 0 ... 1V 10 ... 35 V 9 ... 24 V 0 ... 5 V 14 ... 35 V 12 ... 24 V 0 ... 10 V 19 ... 35 V 16 ...
www.vaisala.com HMDW110 Series Humidity and Temperature Transmitters for High-Accuracy Measurements in HVAC Applications Features/Benefits HMDW110 Series Humidity and Temperature Transmitters for measurements in ducts, outdoor environments, and humid areas. Vaisala HUMICAP® Humidity and Temperature Transmitter Series HMDW110 measure relative humidity and temperature in multiple HVAC applications.
192.5 88 105 HMW110/112 RH+T transmitters for measurements in wet areas Ø3.5 87.5 Ø10 Ø12 Ø4...8 290.8 72 40.6 HMD110/112 RH+T transmitters for ducts 250.2 244.5 105 Ø12 34 26 Ø3.5 Ø12 84 99 100 Ø25 ...60 Ø4...8 207 Ø4...8 HMS110/112 RH+T transmitters for outdoor measurements Ø5.5 Ø4...8 121.5 264.2 43 Humidity 40.6 72 Ø10 Ø3.
Technical Data Models Model number Type Output Special Features Ingress Protection HMW110 Wall-mount, RH+T 2-wire, current output Configurable model* IP65 HMW112 Wall-mount, RH+T 2-wire, current output IP65 HMD110 Duct-mount, RH+T 2-wire, current output Configurable model* IP65 HMD112 Duct-mount, RH+T 2-wire, current output IP65 HMS110 Outdoor, RH+T 2-wire, current output Radiation shield, configurable model* IP65 HMS112 Outdoor, RH+T 2-wire, current output Radiation shield IP65 *Delivered with customer spe
Humidity Find out more at www.vaisala.
www.vaisala.com HMS110 Series Humidity and Temperature Transmitters for High-Accuracy Outdoor Measurements in Building Automation Applications Features/Benefits HMS110 transmitters are equipped with the trusted HUMICAP® 180R – a robust, general-purpose humidity sensor that functions well in high humidity. The sensor’s superior stability ensures long-lasting accuracy and minimal maintenance throughout the transmitter’s lifetime. Easy Installation and Maintenance HMS110 transmitters are easy to install.
Humidity Technical Data Models Model number Type Output Special Features Ingress Protection HMS110 Outdoor, RH+T 2-wire, current output Delivered with customer specific output settings, including calculated humidity parameters and special scaling of outputs. IP65 HMS112 Outdoor, RH+T 2-wire, current output Performance Operating Environment RELATIVE HUMIDITY Measurement range Accuracy Temperature range 0 ... 90 %RH 90 … 100 %RH Temperature range 0 ... 100 %RH +10 ... +30 °C (+50 ...
www.vaisala.com HMDW80 Series Humidity and Temperature Transmitters for Building Automation Applications Features/Benefits HMDW80 Series Humidity and Temperature Transmitters for multiple applications within building automation. Vaisala INTERCAP® Humidity and Temperature Transmitter Series HMDW80 measure relative humidity and temperature in various building automation applications. Transmitters combine easy installation and reliable operation with a low requirement for maintenance.
24 127.3 20 123.7 Ø4.4 Ø4.4 59 32.4 HMW82/83 RH+T and TMW82/83 T-only transmitters for wall-mounting 5.5 60.3 40.6 72 Ø10 HMW88/89(D) RH+T transmitters for measurements in wet areas 192.5 88 105 Ø3.5 Ø3.5 87.5 Ø10 Ø12 Ø4...8 HMD82/83(D) and TMD82/83 RH+T and T-only transmitters for ducts 290.8 72 40.6 250.2 244.5 105 Ø12 34 26 Ø3.5 Ø12 84 99 100 Ø25 ...60 207 Ø4...8 HMS82/83 RH+T transmitters for outdoor measurements Ø5.5 Ø4...8 121.5 264.2 49 Humidity 81 76.
Technical Data Models Model number Type Output Special Features TMW82 Wall-mount, T-only 2-wire, current output TMW83 Wall-mount, T-only 3-wire, voltage output HMW82 Wall-mount, RH+T 2-wire, current output HMW83 Wall-mount, RH+T 3-wire, voltage output HMW88 Wall-mount, RH+T 2-wire, current output Calculated parameters* HMW88D Wall-mount, RH+T 2-wire, current output Display, calculated parameters* HMW89 Wall-mount, RH+T 3-wire, voltage output Calculated parameters* HMW89D Wall-mount, RH+T 3-wire, voltage out
Humidity Specifications for Models HMD82/83D and HMW88/89D Performance Operating temperature range Operating humidity range Mechanics -5 … +60 °C (+23 … +140 °F) 0…100 %RH, non-condensing RELATIVE HUMIDITY Measurement range Accuracy Temperature range 0 ... 90 %RH 90 … 100 %RH Temperature range 0 … 100 %RH +10 ... +30 °C (+50 ... +86 °F) ±3 %RH ±5 %RH -5 ... +10 °C, +30 ... +60 °C (+23 ... +50 °F, +86 ...
www.vaisala.com HMS80 Series Humidity and Temperature Transmitters for Outdoor Measurements in Building Automation Applications Features/Benefits ▪ Reliable outdoor transmitters with integrated radiation shields ▪ ▪ ±3 %RH accuracy ▪ Default output parameters are relative humidity and temperature. Dew point temperature, wet bulb temperature, and enthalpy outputs selectable using dip switches.
Models Humidity Technical Data Operating Environment Model number Type Output Ingress Protection HMS82 HMS83 Outdoor, RH+T Outdoor, RH+T 2-wire, current output IP65 3-wire, voltage output IP65 Operating temperature range Operating humidity range Maximum wind/flow speed Storage temperature Electromagnetic compliance -40 ... +60 °C (-40 ... +140 °F) 0 ... 100 %RH 30 m/s -40 ... +60 °C (-40 ...
www.vaisala.com HMM100 Humidity Module for Environmental Chambers The Vaisala HUMICAP® Humidity Module HMM100. Features/Benefits The Vaisala HUMICAP® Humidity Module HMM100 is an open frame module for integration into environmental chambers. The modules provide a single analog output channel for relative humidity (RH) or dew point (Td). ▪ Full temperature compensation over the operating temperature range of -70 °C ...
Humidity Technical Data Performance Mechanics relative humidity Service cable connector Probe diameter Probe cable lengths Probe material plastics stainless steel Probe mounting clamp Mounting bracket material lid bottom plate Module housing material Measurement range 0 ... 100 %RH Accuracy (incl. non-linearity, hysteresis and repeatability) temperature range -20 ... +40 °C 0 ... 90 %RH ±2 %RH 90 ... 100 %RH ±3 %RH temperature range -40 ... -20 °C, +40 ... +180 °C 0 ... 90 %RH ±2.5 %RH 90 ...
www.vaisala.com Vaisala INTERCAP® Humidity and Temperature Probe HMP60 The HMP60 for extreme conditions. Features/Benefits HMP60 Several Outputs ▪ ▪ ▪ Miniature-size humidity probe There are two configurable voltage outputs with relative humidity, temperature or dew point scaling. Four voltage output ranges are available. ▪ Cable detachable with standard M8 connector The HMP60 is a simple, durable and cost-effective humidity probe.
Operating Environment Measurement range Typical accuracy temperature range 0 ... 90 %RH 90 ... 100 %RH temperature range 0 ... 90 %RH 90 ... 100 %RH Humidity sensor 0 ... 100 %RH 0 ... +40 °C ±3 %RH ±5 %RH -40 ... 0 °C, +40 ... +60 °C ±5 %RH ±7 %RH Vaisala INTERCAP® TEMPERature Measurement range Accuracy over temperature range +10 ... +30 °C -40 ... +10, +30 ... +60 °C -40 ... +60 °C ±0.5 °C ±0.6 °C DEW POINT Measurement range -40 ... +60 °C Typical accuracy temperature range 0 ...
www.vaisala.com Vaisala HUMICAP® Humidity and Temperature Probe HMP110 The HMP110 with excellent stability and high chemical tolerance. Features/Benefits ▪ Miniature-size humidity transmitter ▪ Low power consumption and fast start-up for battery powered applications ▪ Measurement range: 0 ... 100 %RH; -40 ... +80°C ▪ Cable detachable with standard M8 quick connector ▪ Reliable: Latest generation HUMICAP® 180R sensor for best stability and high chemical tolerance. IP65 metal housing.
Performance TEMPERature Measurement range -40 ... +80 °C Accuracy over temperature range 0 ... +40 °C, ±0.2 °C -40 ... 0 °C, +40 ... +80 °C ±0.4 °C Temperature sensor Pt1000 RTD Class F0.1 IEC 60751 DEW POINT Measurement range -40 ... +80 °C Accuracy (incl. non-linearity, hysteresis and repeatability) temperature range 0 ... +40 °C when dew point depression < 15 °C ±1 °C when dew point depression 15 ... 25 °C ±2 °C temperature range -40 ... 0 °C, +40 ...
www.vaisala.com HM70 Hand-Held Humidity and Temperature Meter for Spot-Checking Applications The Vaisala HUMICAP® Hand-Held Humidity and Temperature Meter HM70 is a high-performance, portable humidity reference that is exciting to use.
HMP75, HMP76 and HMP77 Probes Measured Variables MI70 Measurement Indicator Indicator General Relative humidity Menu languages Measurement range 0 ... 100 %RH Accuracy (including non-linearity, hysteresis and repeatability) at +15 ... +25 °C (+59 ... +77 °F) ±1 %RH (0 ... 90 %RH) ±1.7 %RH (90 ... 100 %RH) at -20 ... +40 °C (-4 ... +104 °F) ±(1.0 + 0.008 x reading) %RH at -40 ... +180 °C (-40 ... +356 °F) ±(1.5 + 0.015 x reading) %RH Factory calibration ±0.6 %RH (0 ...
www.vaisala.com Vaisala HUMICAP® Hand-Held Humidity and Temperature Meter HM40 The graph clearly indicates when the readings have stabilized. Benefits The HM40 Humidity Meter can be chosen either in standard or remote probe version. The HM40 is a compact and portable humidity meter at hand everywhere you go. It is an ideal spotchecking tool that is easy to use and provides reliable measurement. It can be used for a wide range of various portable measurement applications.
Performance General Relative Humidity Power-up time <3s Batteries 2 × AA sized, 1.5 V Calculated parameters Td, Tw, a, x, h Menu languages English, German, French, Finnish, Spanish, Swedish, Chinese (simplified), Russian, Japanese, Portuguese Display LCD (140 x 160 pixels) Operation time typical 100 hours (without backlight) Operation temperature range -10 ... +60 °C (+14 ... +140 °F) Storage temperature range -30 ... +70 °C (-22 ...
www.vaisala.com HMI41 Indicator and HMP42/HMP46 Probes The Vaisala HUMICAP® Humidity Indicator HMI41 fitted with the Vaisala HUMICAP® Humidity and Temperature Probes HMP42 or HMP46 can be used for spot checking and field calibration applications. In addition to displaying the humidity and temperature readings, the HMI41 indicator calculates dew point and wet bulb temperature, absolute humidity and mixing ratio. The indicator has an easy-to-read two line liquid crystal display.
HMI41 Indicator Calculated variables HMP46 Probe dew point temperature, absolute humidity, wet bulb temperature, mixing ratio 0.1 %RH; 0.1 °C/°F 4 batteries, type AA (LR 6) Resolution Power supply Battery operation time (alkaline batteries) 72 h continuous use Auto-off function Operating temperature -20 ... +60 °C (-4 ... +140 °F) Storage temperature -40 ... +70 °C (-40 ... +158 °F) Display two line LCD Housing material ABS plastic Housing classification IP53 (with connectors blocked) Weight (incl.
Technical Data General for Probes Dimensions Cable length 1500 mm; extended spiral cable Operating temperature range for electronics -20 ... +60 °C (-4 ...
Humidity Welcome to Vaisala Knowledge eNewsletter SUBSCRIBE Get the latest news about technology, application solutions, instrumentation, and measurement practices. » Subscribe at www.vaisala.com/knowledge Industrial Measurement Spotlight Read about measurements in industrial applications such as compressed air, lubrication, and power production. Topics relate to measurement of relative humidity, dew point, pressure, moisture in oil, and more.
www.vaisala.com SHM40 Structural Humidity Measurement Kit A borehole in concrete and a measurement probe HMP40S inserted in it. Vaisala HUMICAP ® Structural Humidity Measurement Kit SHM40 offers an easy and reliable solution for humidity measurements in concrete and other structures. Concrete dries unevenly and is usually drier on the surface, consequently it is important to measure beneath the surface conditions. The borehole method provides information about the humidity profile under the surface.
HMP40S Probe Relative Humidity Measurement range 0 ... 100 %RH Accuracy (incl. non-linearity, hysteresis, and repeatability): temperature range 0 °C ... +40 °C 0 ... 90 %RH ± 1.5 %RH 90 ... 100 %RH ± 2.5 %RH temperature range -40 °C...0 °C and +40 °C ... +80 °C 0 ... 90 %RH ± 3.0 %RH 90 ... 100 %RH ± 4.0 %RH Factory calibration uncertainty at +20°C (68°F) 0 ... 90 %RH ± 1.1 %RH 90 ... 100 %RH ± 1.8 %RH Humidity sensor HUMICAP® 180R Stability ± 2 %RH over 2 years Temperature Measurement range -40 °C ...
www.vaisala.com HMK15 Humidity Calibrator accepted and reliable method to calibrate humidity instruments. Usually two or three different salt solutions are used. Salts are chosen according to the application. three step plugs to fit all chamber holes (12 mm, 13.5 mm and 18.5 mm) transit covers for secure transportation and long-term storage Certified Salts The HMK15 can be ordered with certified and pre-measured salts.
Humidity Technical Data General Options The standard HMK15 consists of the following parts: Certified and ready dosed salts: Order code: Total uncertainty*: LiCl salt 11 %RH 33 %RH MgCl2 salt NaCl salt 75 %RH 97 %RH K2SO4 salt *Uncertainties given at +20 °C 19729HM 19730HM 19731HM 19732HM (±1.3 %RH) (±1.2 %RH) (±1.5 %RH) (±2.
www.vaisala.com HMP155 Humidity and Temperature Probe Features/Benefits HMP155 with an additional temperature probe and optional Stevenson screen installation kit. The Vaisala HUMICAP® Humidity and Temperature Probe HMP155 provides reliable humidity and temperature measurement. It is designed especially for demanding outdoor applications. Long-term Stability The HMP155 has the proven Vaisala HUMICAP®180R sensor that has excellent stability and withstands well harsh environments.
Technical Data Performance Relative humidity Measurement range 0 ... 100 %RH Accuracy (incl. non-linearity, hysteresis and repeatability) at +15 ... +25 °C (+59 ... +77 ±1 %RH (0 ... 90 %RH) °F) ±1.7 %RH (90 ... 100 %RH) -20 ... +40 °C (-4 ... 104 °F) ±(1.0 + 0.008 x reading) %RH -40 ... -20 °C (-40 ... -4 °F) ±(1.2 + 0.012 x reading) %RH +40 ... +60 °C (+104 ... +140 °F) ±(1.2 + 0.012 x reading) %RH -60 ... -40 °C (-76 ... -40 °F) ±(1.4 + 0.032 x reading) %RH Factory calibration ±0.6 %RH (0 ...
www.vaisala.com DTR500 Solar Radiation and Precipitation Shields 3 Rev. Qty. 2 Description The Vaisala Radiation Shield Series DTR500 are solar radiation and precipitation shields supporting humidity probe installations in outdoor applications. 1 ECO no. - ( DTR504 for HMP60 Design Reviewed D D The maintenance-free DTR500 series shields protect the humidity and temperature sensors from solar radiation and precipitation.
Dimensions Dimensions Dimensions in mm. Dimensions in mm. DTR502B for HMT333, HMT337, HMT363, HMT367, PTU303 and PTU307 DTR503A for HMP155 Humidity Technical Data (ø105) (361) (437) (200) (266) (ø105) (233) (213) DTR502B pole mast installation. DTR503A pole mast installation. DTR502 horizontal beam installation. DTR503 horizontal beam installation.
www.vaisala.com HMT330MIK Meteorological Installation Kit 100 regular probe 95 90 warmed probe 85 80 75 1 2 3 4 5 6 7 After a period of 100 % relative humidity, the warmed probe measures the true humidity, whereas the non-warmed probe takes time to recover from the condensation. and the transmitter calculates the relative humidity from the dew point and temperature values. The HMT337 and PTU307 feature warmed probe technology.
Humidity Technical Data Dimensions Dimensions in mm (inches) 13 Ø220 (8.7) 211 (8.3) 299 (11.8) Ø105 (4.1) Transmitter mounting plate. 371 (14.6) 160 (6.3) 40 (1.6) Ø100 (3.9) Order form 645 (25.4) Pur800 cha(31.5) ser Valid from July 2012 Order no800 . (31.5) Pole mounting plate for support bars, item 10387 Support bar & shield for heated Td probe.
www.vaisala.com HMT300TMK Turbine Mounting Kit for Power Turbine Intake Air Measurement on the humidity sensor and thereby cause errors in measurement. The patented warmed probe prevents condensation from forming on the sensor. Protective Enclosure The HMT300TMK includes a white, painted stainless steel enclosure with an installation kit for the probe. The HMT337 Humidity and Temperature Transmitter is installed in the stainless steel enclosure at the factory, when ordered together with HMT300TMK.
Dewpoint Temperature Outputs Measurement range -40 ... +100 °C (-40 … +212 °F) Accuracy: find the intersection of the dewpoint temperature curve and the dewpoint difference reading (process temperature - dewpoint temperature) on the x-axis and read the accuracy in dewpoint measurement at the y-axis Two standard outputs, third optional Humidity Technical Data 0...20 mA, 4...20 mA, 0 ... 1 V, 0...5 V, 0 ...
Dew Point / TECHNOLOGY DESCRIPTION Vaisala DRYCAP® Sensor for Measuring Dew Point In 1997 Vaisala introduced DRYCAP®, a new type of dew point sensor based on thin-film polymer technology. Since its launch, the DRYCAP product family has grown to encompass a huge range of applications, from drying processes to compressed air and dry chambers. The DRYCAP sensor is particularly renowned for its reliable performance in hot and very dry environments.
Other typical applications include medical gas, dry environments in lithium battery manufacturing, and gas-insulated high-voltage equipment used in the power industry. Upper electrode Vaisala DRYCAP Dew Point Products Thin-film polymer Dew Point Lower electrode Vaisala's dew point instruments are suitable for accurate and stable monitoring of dry conditions in a variety of applications from -80 to +100 °C Td.
www.vaisala.com DMT340 Series Dewpoint and Temperature Transmitters for Very Dry Conditions The display shows measurement trends, real-time data, and measurement history. Features/Benefits ▪ Measures dew point from -60 °C to +80 °C (-76 ... +176 °F) with an accuracy of ±2 °C (±3.
Easy Installation Dew Point DMT340 transmitters are delivered installation-ready, with a variety of installation options to choose from. Dimensions Probe Specifications Dimensions in mm (inches) DMT342 with Small Size Flanged Probe 92.5 (3.64) 41(1.61) 0 ... 50 bar/0 ... 725 psia up to 250 bar/ 3625 psia Probe diameter 71.5 (2.81) 12 mm/0.5" 58.7 (2.31) 36 mm/1.4" Sampling cell HMP302SC 45° 45° M 6 71 (2.
Dimensions DMT347 with Small-Sized Probe Mechanical durability 12 mm/0.5" Installation Fitting body R 3/8" ISO Fitting body G 1/2" ISO Fitting body NPT 1/2" Dimensions in mm (inches) DMT348 with Probe for Pipeline Installations Pressure range Adjustable length 41 ... 149/371 mm/ 1.61 ... 5.87/14.6" 41 (1.61) BALLVALVE-1 Sampling cell DMT242SC or DMT242SC2 R1/2” ISO7/1 NPT 1/2" Ball-valve set Ø 13.5 (0.53) R1/2" ISO 191 (7.52) / 410 (16.1) 41 (1.61) - 149 (5.87) / 41 (1.61) - 379 (14.
Technical Data Inputs and Outputs Dew point Sensor Measurement range For continuous use Accuracy up to 20 bar/290 psia Dew point temperature (°C) 20 ... 50 bar/290 ... 725 psia Vaisala DRYCAP®180M -60 ... +80 °C (-76 ... +176 °F) Td -60 ... +45 °C (-76 ... +113 °F) Td ±2 °C/±3.6 °F (see the accuracy graph below) additional inaccuracy +1 °C Td Area not recommended for continuous measurement Dew point measurement renge lower limit Temperature of measured gas (°C) Dew point accuracy vs.
Dimensions Cable bushing M20x1.5 for cable diameter 8 ... 11mm/0.31 ... 0.43" Conduit fitting 1/2" NPT User cable connector (optional) M12 series 8-pin (male) option 1 female plug with 5 m (16.4 ft.) black cable option 2 female plug with screw terminals USB-RJ45 Serial Connection Cable 219685 Probe cable diameter 5.
You Compress Air. We Make Sure It’s Dry. Dew Point Optimize your compressed air quality with the right dew point instrument Fastest wet-to-dry response time on the market - just minutes Product range from -80°C to +60°C Td with ±2°C accuracy Long, 2-year calibration interval High resistance to contamination – immune to compressor oil Complete recovery from saturated conditions ✔ ✔ ✔ ✔ ✔ For more information, please visit: www.vaisala.com/compressedair www.vaisala.com/compressedair sales@vaisala.
www.vaisala.com DMT345 and DMT346 Dewpoint Transmitters for High Temperature Applications The large graphical display allows the user to check data at a glance. Vaisala DRYCAP ® Dewpoint Transmitters DMT345 and DMT346 are designed to measure and control humidity, especially in dry environments with high temperatures. The Vaisala DRYCAP® Dewpoint Transmitters DMT345 and DMT346 are designed for humidity measurement in industrial drying applications with particularly high temperatures.
DMT346: Reliable in Very Hot Processes The DMT346 provides the best measurement performance at process temperatures between 140 °C and 350 °C. The DMT346 includes a cooling set as standard. The cooling effect can be regulated by adding the cooling fins, or they can be removed from the set for optimal measurement performance. The cooling system has no moving parts, and requires no additional power or cooling utilities, so there is no risk of sensor damage due to mechanical cooling failure.
Technical Data Measured Variables DMT346 Dew point DMT346 Sensor Measurement range Accuracy Vaisala DRYCAP®180S -25 ... +100 °C (-13 ... +212 °F) Td ±2 °C (±3.6 °F) Td See the accuracy graph below Temperature dependence of analog outputs External loads current outputs 0 ... 1V output 0 ... 5V and 0 ... 10V outputs Max. wire size ± 0.005%/°C full scale Inputs and Outputs, DMT345 and DMT346 Mechanics, DMT345 and DMT346 Operating voltage 10 ... 35 VDC, 24 VAC ±20% with optional power supply module 100 .
Technical Data Dimensions Dimensions in mm (inches) DMT345 probe and mounting flange DMT345 and DMT346 transmitter housing 75 (2.96) Dew Point 28 (1.10) 75 (2.95) 12 (0.47) 77 (3.03) 71 (2.80) 13.5 (0.53) 50 (1.97) 41 (1.61) 27 (1.06) 2 (0.08) drilling 16...22 (0.63...0.87) 192 (7.56) 242 (9.53) 183 (7.20) 169 (6.65) DMT346 cooling set 304 (11.97) 164 (6.46) 112.5 (4.43) 40.6 (1.6) 120 (4.72) ø105 (4.13) ø118 (4.65) ø28 (1.1) 8 15 (0.59) (0.31) ø89 (3.5) ø20 (0.79) ø7 (0.28) 96 (3.
www.vaisala.com DMT242 Dewpoint Transmitter for OEM Applications Features/Benefits ▪ ▪ ▪ ▪ ▪ ▪ ▪ Due to its wide measurement range and high long-term stability, the DMT242 is an ideal choice for low dew point industrial applications such as compressed air dryers, plastic dryers and other OEM applications. Vaisala DRYCAP ® The Vaisala DRYCAP® Dewpoint Transmitter DMT242 provides reliable and stable measurements for industrial dryer applications. It is designed for extreme conditions.
Dew Point Temperature Operating Environment Measurement range (typical) -60 ... +60 °C (-76 ... +140 °F) Analog output scalings Option A -80 ...+20 °C (-112 ... +68 °F) Td Option B -60 ... +60 °C (-76 ... +140 °F) Td Option X free scaling (when the dew point is below 0 °C (32 °F), the transmitter outputs frost point) ±2 °C (±3.6 °F) Accuracy with DRYCAP® 180M (see graph below) Temperature higher temperature peaks Relative humidity Pressure Sample flow rate 0 ... +60 °C (32 ... +140 °F) Short-term OK 0 .
www.vaisala.com DMT152 Dewpoint Transmitter for Low Dew Point Measurement in OEM Applications The small and powerful DMT152 measures dew point down to -80 °C.
Technical Data Measured Variables General ±2 °C (3.6 °F) Td ±3 °C (5.4 °F) Td -100 ... -80 °C, -10 ... +20 °C Td (-148 ... -112 °F, +14 ... +68 °F Td) Dewpoint temperature (°C) Analog output scalings option 1 -80 ... +20 °C (-112 ... +68 °F) Td option 2 -100 ... 0 °C (-148 ...
www.vaisala.com DMT143 Miniature Dewpoint Transmitter for OEM Applications Features/Benefits The Vaisala DRYCAP® Dewpoint Transmitter DMT143 is an ideal choice for small compressed air dryers, plastic dryers and other OEM applications. Vaisala DRYCAP® The Vaisala DRYCAP® Dewpoint Transmitter DMT143 is a miniature dew point measurement instrument. The transmitter can be installed directly into pressurized systems at 50 bar (725 psia) maximum pressure.
70 60 50 40 30 20 10 0 -10 -20 -30 -40 -50 -60 -70 -80 -90 Not recommended for continuous measurement Accuracy ±2 °C Td Accuracy ±4 °C Td Accuracy not specified Dewpoint measurement low limit -40 -30 -20 -10 0 10 20 30 40 50 60 Temperature of measured gas (°C) Dew point accuracy vs.
www.vaisala.com DMT132 Dewpoint Transmitter for Refrigerant Dryers Features / Benefits ▪ Affordable dew point transmitter for refrigerant dryers ▪ High accuracy ±1°C (±1.8°F) in the measurement range of refrigerant dryers ▪ Excellent long-term stability resistant to compressor oil and most other chemicals thanks to HUMICAP® technology ▪ Low power requirements, 10 ...
Performance General Dew point Sensor Vaisala HUMICAP®180R Recommended calibration interval 2 years (in refrigerant dryer application) Mechanical connection G 1/2" ISO Operating voltage 10 .... 28 VDC External load max 100 ohm for supply voltages <20 VDC max 500 ohm for supply voltages 20...28 VDC Weight 65 g (2.3 oz) Housing material PPS + 40% GF Housing classification IP65 (NEMA 4) Storage temperature range -40 ... + 80 °C (-40 ...
www.vaisala.com DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications The Vaisala DRYCAP® Hand-Held Dewpoint Meter DM70 offers accurate and fast measurement for industrial dew point applications, such as compressed air, metal treatment and plastics drying. response from ambient to dry conditions exceptionally fast. The DM70 is fitted with the Vaisala DRYCAP® Sensor. The sensor provides reliable, stable and high-performance dew point measurement.
Temperature Measurement range Accuracy at +20 °C (+68 °F) Typical temperature dependence of electronics Temperature sensor -10 ... +60 °C (+14 ... +140 °F) ±0.2 °C (±0.36 °F) ±0.005 °C/°C (±0.005 °F/°F) Pt100 RTD Class F0.1 IEC 60751 other variables available Dew point converted to atmospheric pressure, ppm volume and ppm weight concentration, absolute humidity, mixing ratio, relative humidity Measured Variables, DMP74B and DMP74C (for SF6 gas) Probes Dew point Measurement range (typical) -70 ...
www.vaisala.com DSS70A Portable Sampling System and Sampling Cells for DM70 The DSS70A provides a compact solution for field checking dew point where direct measurement is difficult. Typical applications for the sampling system are metal treatment and plastics drying processes. DSS70A Portable Sampling System The DSS70A is designed to provide dew point sampling flexibility for the DM70 hand-held dew point meter.
DMCOIL DMT242SC pressure limit DMT242SC2 DSS70A Sampling System and DSC74B/C Sampling Cells Filter Sample Gas Inlet Sampling Cell for Td Probe Needle Valve Flow meter Sample Pump Sample Gas Outlet pressure limit Material for all sampling cells DSS70A The DSS70A sampling system includes a filter to clean the dirty sample gas and a needle valve to control the sample flow rate with the flow meter. A sample pump is used to generate a sample flow from processes at ambient pressure.
www.vaisala.com DPT146 Dewpoint and Pressure Transmitter for Compressed Air Features/Benefits Dimensions Dimensions in mm (inches) 104 128.5 120.5 ▪ ▪ Proven sensor technology ▪ ▪ Pressure: 1 … 12 bar ▪ Digital output RS-485 with MODBUS Compatible with the Vaisala Hand-Held DM70 for easy spot checking, local display and data logging Dew point: -60 … +30 °C (-76 ... +86 °F) Tdf with accuracy of ±2°C (±3.
Technical Data Measured Parameters -60 ... +30 °C (-76 ... +86 °F) 1 ... 12 bar (14.5 ... 174 psi) -40 ... +80 °C (-40 ... +176 °F) Calculated Parameters ppm moisture, by volume Dew point, converted to atmospheric pressure 1 ... 40 000 ppm -75 … +30 °C (-103 ... +86 °F) Performance Dew point accuracy ±2 °C (±3.6 °F) Pressure accuracy at 23 °C (73.4 °F) ±0.4 %FS Pressure temperature dependence ±0.01 bar / 10 °C (18 °F) Temperature accuracy 0 ... 40 °C (+32 ... +104 °F) ±0.5 °C (± 0.9 °F) -40...
www.vaisala.com DPT145 Multiparameter Transmitter for SF6 Gas Features/Benefits The Vaisala Multiparameter Transmitter DPT145 with the DILO DN20 connector. The Vaisala Multiparameter Transmitter DPT145 for SF6 Gas is a unique innovation that enables online measurement of dew point, pressure, and temperature. It also calculates four other values, including SF6 density. The DPT145 is especially well suited for integration into OEM systems. board, from investment to installation, operation and servicing.
Technical Data Operating Environment Operating temperature of electronics -40 … +60 °C (-40 ... +140 °F) Operating Pressure 0 … 50 bar (0…725 psi) Relative humidity 0...100 % Measured gases SF6, SF6/N2 mixture Dewpoint Pressure, absolute Temperature -50 ... +30 °C (-58 ... +86 °F) 1 ... 12 bar (14.5 ... 174 psi) -40 ... +80 °C (-40 ... +176 °F) 1 ... 12 bar (14.5…174 psi) 0 ... 100 kg/m3 40 ... 40 000 ppm -65 … +30 °C (-85 ... +86 °F) Performance Dewpoint accuracy ±3 °C (±5.
www.vaisala.com Dimensional Drawings of the DPT145 Multiparameter Transmitter 29.5 37 M30x2 AW27 20 52 8 44 80 44 7 53.5 63.5 200 41.6 The DPT145 with the ABB Malmkvist connector. 29.5 30 The DPT145 with the weather shield. AW27 AW36 M45x2 G1/2 ISO 228-1 6 30.6 AW 32 51 59 29.5 4.8 AW8 145.5 7 20 83 7 160 60 70 AW50 33 The DPT145 with the DILO DN20 connector. 57.7 20 5 152 The DPT145 with the Alstom connector. Ø 20 Ø 29.5 AW32 / M26x1.5 7 AW 32 35.
Dew Point Outages Happen 24/7. So Should Monitoring.
Moisture in Oil / TECHNOLOGY DESCRIPTION Vaisala HUMICAP® Sensor for Measuring Moisture in Oil Water is a common contaminant in industrial oils. Water contamination deteriorates the performance of the oil, be it used for lubrication, cooling, insulation or other purposes. High moisture content increases the risk of corrosion, overheating, machine malfunction and other problems and can ultimately lead to costly failure and unscheduled downtime.
In contrast to traditional measurement techniques, water activity measurement is independent of oil type. Regardless of the saturation point of the fluid, water activity measurement always provides a true indication for the risk of free water formation, even when the saturation point is increasing or decreasing. In its simplicity, water activity value is understandable at a glance. Trends can be quickly identified.
www.vaisala.com MMT330 Moisture and Temperature Transmitter Series for Oil The display shows measurement trends, real-time data, and measurement history. The MMT330 transmitter family offers reliable performance for the demanding measurement of moisture in oil.
Water Content as ppm Conversion The display alarm allows any measured parameter to be tracked, with freely configurable low and high limits. In addition to water activity, the MMT330 can output ppm, the average mass concentration of water in oil. Vaisala has this conversion readily available for mineral transformer oil. Versatile Outputs and Data Collection For serial interface the USB connection, RS232, and RS485 can be used.
Dimensions Installation Options Dimensions in mm (inches) 12 mm / 0.5" Installation 35 (1.38) Flange 36 mm / 1.4" 65.5 (2.58) -40 ... +180 °C (-40 ... 356 °F) Dimensions Pressure range Dimensions in mm (inches) ISO 3/8" or NPT1/2" MMT337 with Small-Sized Probe 0 ... 10 bar / 0 ... 145 psia 12 mm / 0.5" Installation Fitting body R 3/8" ISO Fitting body 1/2" ISO Fitting body NPT 1/2" Hex 22m (0.87) Ø 12 (0.47) Probe diameter 33 (1.30) 77 (3.03) Temperature -40 ... +180 °C (-40 ...
Measured Values Water activity Measurement range aw 0 ... 1 Accuracy (including non-linearity, hysteresis and repeatability) 0 ... 0.9 ±0.02 0.9 ... 1.0 ±0.03 Response time (90%) at +20 °C in still oil (with stainless steel filter) 10 min. Sensor HUMICAP® 180L2 Performance Temperature Measurement range MMT332 MMT337 MMT338 Accuracy at +20 °C (+68 °F) -40 ... +180 °C (-40 ... +356 °F) -40 ... +180 °C (-40 ... +356 °F) -40 ... +180 °C (-40 ... +356 °F) ± 0.2 °C (0.
www.vaisala.com MMT310 Series Moisture and Temperature Transmitter for Oil Water Content as PPM Calculation for Transformer Oils PPM units are traditionally used in transformer applications. They indicate the average mass concentration of water in oil. The ppm calculation for mineral oil based transformer oil is optional in the MMT310 series. Diverse Applications and Demanding Conditions The MMT310 series can be used in lubrication and hydraulic systems as well as in transformers.
Technical Data General Water activity Measurement range aw Accuracy (including non-linearity, hysteresis and repeatability) 0 ... 0.9 0.9 ... 1.0 Response time (90 %) at +20 ˚C in still oil (with stainless steel filter) Sensor 0 ... 1 ±0.02 ±0.03 10 min Vaisala HUMICAP® Temperature Measurement range Typical accuracy at +20 ˚C Typical temperature dependence of electronics Sensor -40 ... +180 ˚C (-40 ... +356 ˚F) ±0.1 ˚C (±0.18 ˚F) ±0.05˚C/ ˚C (±0.005 ˚F/˚F) Pt100 RTD Class F0.
www.vaisala.com MMT162 Compact Moisture in Oil and Temperature Transmitter for OEM Applications Water Activity Measurement The MMT162 measures moisture in oil in terms of the water activity (aw) and temperature (T). Water activity directly indicates whether there is a risk of free water formation. The measurement is independent of oil type, age and temperature. The ppm calculation for mineral oil based transformer oil is optional in the MMT162.
Measured Values General Water activity Sensor HUMICAP® Cable connections (2 ports) M8, 4 pin Minimum operating voltage with RS-485 output 14 ... 28 VDC voltage output 16 ... 28 VDC current output 22 ... 28 VDC Supply current normal measurement 20 mA + load current External load for voltage output min. 10 kOhm current output max.
www.vaisala.com MM70 Hand-Held Moisture and Temperature Meter for Spot-Checking in Oil Numerical and Graphical Display The MM70 features a multilingual, menu-based user interface and a backlit LCD display. The measurement parameters can be numerically and graphically displayed and logged into the meter's memory at the same time. An analog output option is also available. The MM70 is an ideal tool for the preventive maintenance of oil-filled systems.
Technical Data Performance Display Measurement range aw 0 ... 1 Accuracy (including nonlinearity, hysteresis and repeatability) When calibrated against salt solutions (ASTM E104-85): 0 ... 0.9 ±0.02 0.9 ... 1.0 ±0.03 Maximum achievable accuracy when calibrated against high-quality, certified humidity standards: 0 ... 0.9 ±0.01 0.9 ... 1.0 ±0.02 Response time (90%) at +20 °C (+68 °F) in still oil (with stainless steel filter) 10 min.
Carbon Dioxide / TECHNOLOGY DESCRIPTION Vaisala CARBOCAP® Sensor for Measuring Carbon Dioxide First launched in 1997, the Vaisala CARBOCAP® carbon dioxide (CO2) sensor features a groundbreaking innovation – the micromachined, electrically tunable Fabry-Perot Interferometer (FPI) filter for built-in reference measurement.
Vaisala CARBOCAP instruments are well suited to a wide range of applications, from ppm (parts per million) to percentage-level CO2 measurements. Since CO2 is harmful in high concentrations, it is present at percentage levels only within closed processes such as fermentation and controlled-atmosphere storage environments. Percentage-level measurements are also typical in life-science applications such as CO2 incubation.
www.vaisala.
Technical Data Pressure Inputs and Outputs Effect on accuracy with pressure compensation: 0 ... 1000 ppm 0 ... 2000 - 2 % CO2 range options Pressure (hPa) Accuracy (% of reading) 900 ... 1050 ±0.5 ±1 700 ... 1300 ±1 ±2 Integrated pressure sensor is not included in GMP343 Operating voltage Power consumption without optics heating with optics heating Long term stability easy moderate harsh * Always at least ±10 ppm CO2.
www.vaisala.com Dimensional Drawings of the GMP343 Carbon Dioxide Probe Dimensions in mm (inches) 1.77 45 6.88 175 7.4 188 7.13 181 7.64 194 0.94 24 2.07 52.5 2.17 55 2.17 55 Diffusion model Flow-through model Please contact us at www.vaisala.com/requestinfo www.vaisala.com 126 Ref. B211239EN-A ©Vaisala 2012 Scan the code for more information This material is subject to copyright protection, with all copyrights retained by Vaisala and its individual partners. All rights reserved.
Carbon Dioxide Carbon Dioxide Measurement with Superior Stability and Ability. Unique built-in reference measurement and robust reliability. Stability is what everyone promises. But it is exactly what Vaisala CARBOCAP® successfully delivers with its unique built-in reference measurement. Vaisala CARBOCAP® combines the ease and cost efficiency of extra long calibration intervals with leading accuracy on both ppm and percentage level measurements.
www.vaisala.com GMT220 Series Carbon Dioxide Transmitters for Industrial Applications The GMT220 transmitters withstand harsh and humid environments.
Technical Data Operating Environment 0 ... 2 % 0 ... 3 % 0 ... 5 % 0 ... 10 % 0 ... 20 % GMT222 for low concentrations 0 ... 2000 ppm 0 ... 3000 ppm 0 ... 5000 ppm 0 ... 7000 ppm 0 ... 10 000 ppm Accuracy (including repeatability, non-linearity and calibration uncertainty) at 25 °C and 1013 hPa GMT221 ±(1.5 % of range + 2 % of reading) (applies for concentrations above 2 % of full scale) GMT222 ±(1.5 % of range + 2 % of reading) Temperature dependence, typical -0.
www.vaisala.com GMM220 Carbon Dioxide Modules for Demanding OEM Applications The Vaisala CARBOCAP® Carbon Dioxide Module Series GMM220 withstand harsh conditions. They provide high carbon dioxide measurement accuracy over wide temperature and relative humidity ranges.
Technical Data Carbon Dioxide Mechanics Measurement ranges GMM221 for high concentrations Probe housing material PC plastic Housing classification (probe only) IP65 Weight: GMM221 (w/2m cable) max. 180 g Weight: GMM222 (w/2m cable) max. 200 g Probe cable length 0.6 m, 1 m (high temperature cable), 2 m, 6 m or 10 m (optional) Inputs and Outputs Outputs 0 ... 20 or 4 ... 20 mA, 0 ... 1 V, 0 ... 2V, 0 ... 2.5 V, or 0 ...
www.vaisala.com GMM111 Carbon Dioxide Module The Vaisala CARBOCAP® Carbon Dioxide Module GMM111 is a CO2 measurement module with flow-through aspiration. Features/Benefits ▪ Compact CO2 module with flow-through aspiration ▪ Ideal for control of CO2 concentrations in incubators ▪ Incorporates Vaisala CARBOCAP®, the silicon based NDIR sensor with unique internal referencing ▪ Advanced single-beam, dual wavelength measurement with no moving parts ▪ Measurement range options 0 ... 5 %, 0 ...
Technical Data Dimensions CO2 measurement range 0 ... 5 %, 0 ... 10 % or 0 ... 20 % Accuracy (including repeatability, non-linearity and calibration uncertainty) ±(1.5% of range + 3 % of reading) Long-term stability ±0.5 %CO2/year 0 ... 8 %CO2 ±1 %CO2/year 8 ... 12 %CO2 ±2 %CO2/year 12 ... 20 %CO2 < 1 min, when flow > 0.2 l/min Response time T90 Flow rate dependence < 1 l/min flow no effect 1 ... 10 l/min flow 4 % of reading/ l/min Temperature dependence, -0.
www.vaisala.com GM70 Hand-Held Carbon Dioxide Meter for Spot-Checking Applications accuracy is not affected by dust, water vapor or most chemicals. The GM70 has a two-year recommended calibration interval. The Vaisala CARBOCAP® Hand-Held Carbon Dioxide Meter GM70 is the demanding professional’s choice for hand-held carbon dioxide measurement. The meter consists of the indicator (center) and probe, used either with the handle (left) or pump (right).
CO2 Volume Concentration Measurement MI70 Indicator General Measurement ranges High concentrations short probe (GMP221) Menu languages 0 ... 2 % 0 ... 3 % 0 ... 5 %, 0 ... 10 %, 0 ... 20 % Low concentrations 0 ... 2000 ppm long probe (GMP222) 0 ... 3000 ppm, 0 ... 5000 ppm, 0 ... 7000 ppm, 0 ... 10,000 ppm Accuracy (including repeatability, non-linearity and calibration uncertainty) at 25 °C and 1013 hPa GMP221 ±(1.
www.vaisala.com GMW90 Series Carbon Dioxide, Temperature, and Humidity Transmitters for Demand Controlled Ventilation Applications Features/Benefits GMW90 Series Carbon Dioxide, Temperature and Humidity Transmitters for HVAC are available with either a display opening or a solid front. An optional traffic light indication can also be selected.
Technical Data GMW93 GMW93D GMW94 GMW94D GMW93R GMW93RD GMW93RA CO2+T CO2+T CO2+T CO2+T CO2+T+RH CO2+T+RH CO2+T+RH Stability in typical HVAC applications Humidity sensor ±0.
Dimensions Dimensions in mm 59.5 30 29.8 4.4 30.5 81 Please contact us at www.vaisala.com/requestinfo www.vaisala.com 138 34 33.5 9 132.7 27 47.5 5 Ø Ref. B211296EN-B ©Vaisala 2014 Scan the code for more information This material is subject to copyright protection, with all copyrights retained by Vaisala and its individual partners. All rights reserved. Any logos and/or product names are trademarks of Vaisala or its individual partners.
Decorative Cover for Vaisala GMW90 and HMW90 Series Transmitters Carbon Dioxide The decorative cover helps to camouflage the transmitter to the surroundings. Transmitter location can be chosen with optimal measurement in mind without compromising aesthetics. www.vaisala.com/GMW90 www.vaisala.
www.vaisala.com GMW116 Carbon Dioxide and Temperature Transmitter for Ventilation Control The Vaisala CARBOCAP® Carbon Dioxide and Temperature Transmitter GMW116. Features/Benefits ▪ ▪ Compact dual-parameter transmitter: measures both CO2 and T Incorporates Vaisala CARBOCAP®, the silicon based NDIR sensor with unique internal referencing The Vaisala CARBOCAP® Carbon Dioxide and Temperature Transmitter GMW116 is the compact-size transmitter that measures both carbon dioxide and temperature.
Technical Data Performance Dimensions CO2-measurement range Temperature measurement range Accuracy (including repeatability, nonlinearity and calibration uncertainty) Long-term stability Response time T90 Temperature dependence Pressure dependence, typical Temperature measurement accuracy Warm-up time Product lifetime 0 ... 2000 ppm 0 ... +50 °C ± (2 % of range + 2 % of reading ± 5 % of range/5 years 1 min compensated +0.15 % of reading/hPa ± 0.7 °C at 25 ° 1 min, 10 min for full.
www.vaisala.
Technical Data Dimensions 66.3 60 .2 120.9 60 66.7 Ø4 124 R2.1 84 R3.5 Temperature -5 ... +45 °C (23 ... 113 °F) Humidity 0 ... 85 %RH Pressure 700 ... 1200 hPa Electromagnectic compatibility Complies with EMS standard EN61326-1, Generic Environment 25.8 Operating Environment Carbon Dioxide Product lifetime Dimensions in mm 85 .
www.vaisala.com GMD/W20 Series Carbon Dioxide Transmitters for Demand Controlled Ventilation Applications affected by dust, water vapor and most chemicals. Versatile Transmitters The GMD/W20 series transmitters can be used independently, or incorporated into building energy management systems. The series consists of duct mount units GMD20/D and wall mount units GMW21/D and GMW22/D. Version D has a display. The two sizes in the wall-mount series make them easy to install for most standard junction boxes.
Technical Data Performance Accessories and Options Carbon dioxide measurement Display and relay option for GMD/W21/22 Relay output option LonWorks® module with CO2 signal (Not available when display option is added) LonWorks® module with both CO2 signal and temperature signals (Not available when display option is added) Serial COM adapter Analog temperature module for GMW21 (not available when display option is added) Hand-held meter for field verification Measurement range 0 ...
www.vaisala.
Technical Data Dimensions Operating Environment Inputs and Outputs Operating voltage Power consumption Outputs Ø3.2 24 V (±20 %) AC/DC <2 W 4 … 20 mA, 0 … 10 V, RS-485, 2-wire, non-isolated 30 Please contact us at www.vaisala.com/requestinfo www.vaisala.com mA V 0 B A 0 24V 6 Temperature -5 ... +45 °C (23 ... 113 °F) Humidity 0 ... 85 %RH Pressure 700 ...
Pressure / TECHNOLOGY DESCRIPTION Vaisala BAROCAP® Sensor for Measuring Pressure First introduced in 1985, Vaisala BAROCAP® is a silicon-based micromechanical pressure sensor that offers reliable performance in a wide variety of applications, from meteorology to industrial measurements. Combining two powerful technologies – single-crystal silicon material and capacitive measurement – BAROCAP sensors feature low hysteresis combined with excellent accuracy and long-term stability.
BAROCAP Applications for Measurements in Pressurized Systems Vaisala DRYCAP® dew point instruments have a long history of providing reliable and stable measurements in compressed air systems and SF6 insulation gas monitoring in high-voltage equipment. In addition to the need for dew point measurement, these two applications also share the need for accurate and stable pressure measurement.
www.vaisala.com PTU300 Combined Pressure, Humidity and Temperature Transmitter for Demanding Applications The Vaisala Combined Pressure, Humidity and Temperature Transmitter PTU300 is a versatile, multi-purpose instrument.
Versatile Outputs and Data Collection The PTU300 comes with a standard RS232 serial interface. The output format is compatible with major GPS receivers and NMEA-coded messages. An isolated RS485 is available as an option. The PTU300 is also capable of applying the MODBUS communication protocol and, together with an appropriate connection option, provides either MODBUS RTU (RS485) or MODBUS TCP/IP (Ethernet) communication.
PTU300 Models Dimensions M9x1 Ø 12 (0.47) Dimensions in mm (inches) 37.5 (1.48) 78.5 (3.09) PTU303 probe for outdoor use 98.5 (3.88) Dimensions M9x1 Ø 12 (0.47) Dimensions in mm (inches) 37.5 (1.48) 79.5 (3.13) PTU307 warmed probe for demanding meteorological installations Ø 6 (0.24) Ø 3.2 (0.13) 99.5 (3.92) 130 (5.12) Dimensions Ø 6 (0.24) Ø 3.2 (0.13) Dimensions in mm (inches) 130 (5.
Technical Data Performance Barometric pressure 500 ... 1100 hPa, 50 ... 1100 hPa 500 ... 1100 hPa 500 ... 1100 hPa 50 ... 1100 hPa Class A Class B Linearity ±0.05 hPa ±0.10 hPa ±0.20 hPa Hysteresis* ±0.03 hPa ±0.03 hPa ±0.08 hPa Repeatability* ±0.03 hPa ±0.03 hPa ±0.08 hPa Calibration ±0.07 hPa ±0.15 hPa ±0.20 hPa uncertainty** Accuracy at +20 °C*** ±0.10 hPa ±0.20 hPa ±0.30 hPa Temperature ±0.1 hPa ±0.1 hPa ±0.3 hPa dependence**** Total accuracy ±0.15 hPa ±0.25 hPa ±0.45 hPa (-40 ... +60 °C/ -40 ...
Probe cable diameter PTU303 6.0 mm other probes 5.5 mm Standard probe cable lengths 2 m, 5 m or 10 m (Additional cable lengths available, please see order form for details) Housing material G-AlSi 10 Mg (DIN 1725) Housing classification IP 66 IP65 (NEMA4X) with local display Weight depending on selected probe 1.0 - 3.
Products, spare parts and accessories Pressure Convenient product comparison tool Invoice option (30 days NET) store.vaisala.
www.vaisala.com PTB330 Digital Barometer for Professional Meteorology, Aviation, and Industrial Users against one another and provides information on whether these are within the set internal difference criteria. This unique feature provides redundancy in pressure measurement. Thus, users also get a stable and reliable pressure reading at all times as well as a pre-indication of when to service or recalibrate the barometer. QNH and QFE Vaisala BAROCAP® Digital Barometer PTB330 with a new trend display.
Technical Data Inputs and Outputs Barometric pressure range 500 ... 1100 hPa Class A ±0.05 hPa ±0.03 hPa ±0.03 hPa ±0.07 hPa ±0.10 hPa Linearity* Hysteresis* Repeatability* Calibration uncertainty** Accuracy at +20 °C (+68 °F) *** Class B ±0.10 hPa ±0.03 hPa ±0.03 hPa ±0.15 hPa ±0.20 hPa Barometric pressure range 50 ... 1100 hPa Class B ±0.20 hPa ±0.08 hPa ±0.08 hPa ±0.15 hPa ±0.
www.vaisala.com Dimensional Drawings of the PTB330 Digital Barometer Dimensions Dimensions in mm (inches) 183 (7.20) 169 (6.65) 96 (3.78) ∅7 (0.28) 77 (3.03) 116 (4.57) 71 (2.80) Mounting Options Mounting with Wall Mounting Kit Mounting with DIN Rail Installation Kit Pole Installation with Installation Kit for Pole or Pipeline Mounting Rain Shield with Installation Kit HUMICAP® is a registered trademark of Vaisala. Please contact us at www.vaisala.com/requestinfo www.vaisala.com 158 Ref.
Welcome to Vaisala Knowledge eNewsletter SUBSCRIBE Get the latest news about technology, application solutions, instrumentation, and measurement practices. » Subscribe at www.vaisala.com/knowledge Industrial Measurement Spotlight Read about measurements in industrial applications such as compressed air, lubrication, and power production. Topics relate to measurement of relative humidity, dew point, pressure, moisture in oil, and more.
www.vaisala.com Take the Lab to the Field Vaisala Barometric Pressure Transfer Standard PTB330TS Operational PTB330TS unit includes a PTB330 digital barometer, hand-held MI70 indicator, optional HMP155 humidity and temperature probe, optional MI70 Link PC software, a user’s guide and a weatherproof transfer case equipped with a shoulder strap. PTB330TS transport case Barometric Pressure Transfer Standard PTB330TS makes field inspection accurate and easy.
The PTB330 Digital Barometer The HMP155 probe The MI70 hand-held indicator displaying the prevailing pressure in hPa PTB330 Digital Barometer against one another and provides information on whether these are within the set internal difference criteria. This unique feature provides redundancy in pressure measurement. Thus, users get a stable and reliable pressure reading at all times as well as a pre-indication of when to service or re-calibrate the barometer.
www.vaisala.com Vaisala Barometric Pressure Transfer Standard PTB330TS Technical Data These specifications apply when MI70, PTB330 and HMP155 are used together in the PTB330TS product. For individual specifications, please refer to the product documentation and brochures of the PTB330 and HMP155. General Operating temperature range Operating humidity range Maximum pressure limit Power supply -10 ... +40 °C (+14 ... +104 )°F non-condensing 5000 hPa abs.
Technical Data Available Parameters Accessories Pressure parameters Humidity and temperature parameters P, P3h, HCP, QFE, QNH RH, T, Tdf, Td, x, Tw Inputs and Outputs PTB330 supply voltage PTB330 data interface PTB330 serial I/O connectors HMP155 data interface HMP155 serial I/O connector 2 RS-232 (accessible only with MI70 Link software) 10 ...
www.vaisala.com PTB210 Digital Barometer The PTB210 paired with the SPH10 Static Pressure Head. The Vaisala BAROCAP® Digital Barometer PTB210 is a reliable outdoor barometer that withstands harsh conditions. For Harsh Environments The Vaisala BAROCAP® Digital Barometer PTB210 is ideal for outdoor installations and harsh environments. The barometers are designed to operate in a wide temperature range, and the electronics housing provides IP65 (NEMA 4) standardized protection against sprayed water.
Technical Data Pressure range (order specified) serial output analog output Operating temperature range Humidity range 500 ... 1100 hPa 50 ... 1100 hPa 500 ... 1100 hPa 600 ... 1060 hPa 800 ... 1060 hPa 900 ... 1100 hPa -40 ... +60 °C (-40 ... +140 °F) non-condensing Accuracy Serial output (units in hPa) Pressure range Non linearity* Hysteresis* Repeatability* Calibration uncertainty** Accuracy at +20 °C (+68 °F)*** Temperature dependence**** Total accuracy*** -40 ... +60 °C (-40 ...
www.vaisala.com PTB110 Barometer for Industrial Use The Vaisala BAROCAP® Barometer PTB110 offers outstanding long-term stability. Features/Benefits ▪ ▪ ▪ ▪ ▪ Vaisala BAROCAP® sensor ▪ Output voltage 0 ... 2.5 or 0 ... 5 VDC ▪ Current consumption less than 4 mA ▪ Mountable on a (35 mm wide) DIN rail ▪ NIST traceable (certificate included) 166 Several pressure ranges Accuracy ±0.
Technical Data Operating Range (1 hPa=1mbar) Accuracy Pressure ranges Linearity* ±0.25 hPa Hysteresis* ±0.03 hPa Repeatability* ±0.03 hPa Pressure calibration uncertainty** ±0.15 hPa Voltage calibration uncertainty ± 0.7 mV Frequency calibration uncertainty ± 0.3 Hz Accuracy at +20 °C*** ±0.3 hPa * Defined as ±2 standard deviation limits of end-point non-linearity, hysteresis error or repeatability error.
www.vaisala.com SPH10/20 Static Pressure Heads for Minimizing Wind Induced Error Wind induced effects are one of the main sources of error when measuring barometric pressure. Variations due to strong and gusty wind can be overcome by using a static pressure head to ‘filter out’ the effect of dynamic pressure. The Vaisala Static Pressure Head Series SPH10/20 are designed to minimize the errors caused by wind. The wind tunnel tested structure is both horizontally and vertically symmetrical.
Technical Data General Operating temperature Weight SPH10 SPH20 Material Mounting Hose connection SPH20 -60 ... +80 °C (-76 ... +176 °F) 800 g 1360 g PC plastic, offshore aluminium with 2 bolts (M6 X 20 mm min) barbed fitting for 4 mm I.D. hose or Rp1/4 thread (parallel) Electrical connections Power supply M12 connector factory setting 12 V changed connection 24 V Thermostat switching temperature On Off Power consumption during heating +4 °C (±3 °C) +39.2 °F (±4.4 °F) +13 °C (±3 °C) +55.4 °F (±4.
www.vaisala.com Vaisala Differential Pressure Transmitter PDT102 Vaisala Differential Pressure Transmitter PDT102 with process valve actuator and test jacks. Features ▪ ▪ ▪ ▪ ▪ ▪ 170 In-place system calibration and on-line monitoring without disturbing process tubes with optional process valve actuator and test jacks Ultrathin profile ideally suited for DIN rail mount reduces installation and calibration costs High accuracy, two options; 0.25% or 0.
Technical Data Performance Measurement ranges (bidirectional) ±50 Pa ±0.25 in H2O Overpressure proof pressure 0.7 bar burst pressure 1.7 bar static pressure 1.7 bar Pressure type differential, gauge, vacuum and compound Accuracy (incl. non-linearity, hysteresis, repeatability and 0.25 % span or 0.5 % span, zero/span calibration settings) depending on choice Repeatability for 0.25 % span accuracy 0.03 % for 0.5 % span accuracy 0.05 % Electrical resolution 1 x 10-4 span Long-term stability ≤0.
www.vaisala.com Vaisala Differential Pressure Transmitter PDT101 Vaisala Differential Pressure Transmitter PDT101 with current output (black) and voltage output (green).
Technical Data Performance Measurement ranges (bidirectional) ±60 Pa ±0.25 in H2O Overpressure proof pressure 1.0 bar burst pressure 1.7 bar static pressure 1.7 bar Pressure type differential, gauge, vacuum and compound Accuracy (incl. non-linearity, hysteresis, repeatability and zero/span calibration settings) 0.4 % span Long-term stability ≤0.5 % span/year Response time (10 ... 90 %) 250 ms Warm-up time 15 s Compensated temperature range +2 ... +54 °C (+35.6 ... +129.2 °F) Temperature dependence ±(0.
Weather / TECHNOLOGY DESCRIPTION Vaisala Wind and Weather Sensor Technologies for Measurements in Industrial Applications Vaisala's long history in wind and weather measurements started already in the 1930s from the development of a radiosonde to measure the conditions in the upper atmosphere. Today, Vaisala wind and weather instruments are used in dozens of applications and industries all over the world.
Vaisala Sensor Technologies for Wind and Rain Measurements Vaisala WINDCAP® Sensor Vaisala RAINCAP® Sensor Vaisala WINDCAP Ultrasonic Wind Sensor uses ultrasound to determine wind speed and direction. The sensor has no moving parts, which makes it independent of the limitations of mechanical wind sensors such as friction, inertia, time constant, over-speeding, and starting treshold.
www.vaisala.com Vaisala Weather Transmitter WXT520 Access to Real Time Weather Data Features/Benefits The WXT520 has an automatic control circuit that switches the heating on at low temperatures. To measure wind speed and direction, the WXT520 has the Vaisala WINDCAP® Sensor that uses ultrasound to determine horizontal wind speed and direction. The array of three equally spaced transducers on a horizontal plane is a Vaisala specific design.
Technical Data Wind Air Temperature Speed 0 ... 60 m/s 250 ms ±3% at 10m/s 0.1 m/s, 0.1km/h, 0.1 mph, 0.1 knots Range -52 ... +60 °C (-60 ... +140 °F) Accuracy for sensor at +20 °C ±0.3 °C (±0.5 °F) Accuracy over temperature range (see graph below) Δ °C 0.7 0.6 0.5 0.4 Direction azimuth response time accuracy output resolution and unit 0.3 0.2 0 ... 360° 250 ms ±3° 1° 0.1 °C 0 -0.1 -0.2 -0.3 -0.4 -0.5 -0.6 -0.
www.vaisala.com WMT52 Ultrasonic Wind Sensor Applications The WMT52 is ideal for use in marine applications as the housing with the mounting kit is water resistant. The WMT52 is also suitable for wind energy and environmental monitoring, for example, for measuring the distribution of air pollution and road tunnel ventilation. Easy to Install The WMT52 is delivered fully assembled and configured from the factory.
Technical Data Wind Power Supply Wind speed Range Response time Available variables Accuracy Output resolution 0 ... 60 m/s 250 ms average, maximum, and minimum ±3% at 10m/s 0.1 m/s (km/h, mph, knots) Operating voltage Power consumption on average minimum maximum typical 5 ... 32 VDC 0.1 mA at 12 VDC 14 mA at 5 VDC 3 mA at 12 VDC (default measuring intervals) 5 ... 32 VDC / 5 ... 30 VACRMS) Heating voltage Wind direction Azimuth Response time Available variables Accuracy Output resolution 0 ...
www.vaisala.com WA15 Wind Set for High Performance Wind Measurement The output pulse rate is directly proportional to wind speed (e.g. 246 Hz = 24.6 m/s). However, for the highest accuracy, the characteristic transfer function should be used to compensate for starting inertia. Sensitive Wind Vane The WAV151 is a counter-balanced, low-threshold, optoelectronic wind vane. Infrared LEDs and phototransistors are mounted on six orbits on each side of a 6-bit GRAY-coded disc.
Technical Data Vaisala Anemometer WAA151 Vaisala Wind Vane WAV151 Wind Speed Wind Direction Measurement range 0.4 ... 75 m/s Starting threshold < 0.5 m/s * Distance constant 2.0 m Characteristic transfer function U = 0.328 + 0.101 × R (where U = wind speed [m/s], R = output pulse rate [Hz]) Accuracy (within range 0.4 ... 60 m/s) with characteristic transfer function ± 0.17 m/s ** with transfer function U = 0.1 × R ± 0.5 m/s*** Measurement range at wind speed 0.4 ...
Vaisala Service Offering Confidence on Day 1, Confidence for Years Calibration Maintains Reliability Regular calibration ensures your high-precision instruments continue to provide accurate, high-quality data. Choosing the right calibration laboratory can be as important for lifetime accuracy and reliability as the initial selection of the instrument. Identifying and setting calibration intervals are based on the appliction, product usage and the company's quality system.
Calibration Maintains and Documents Accuracy Vaisala offers both standard service center calibrations and calibration services that have been accredited by the world’s leading accreditation authorities. Accrediting bodies audit management, quality systems and technical competence and Vancouver, Canada regulating bodies or to comply with a company’s quality standards.
Vaisala Filters Group A: 12 mm Diameter, Standard Length, Female Threads 10159HM Membrane filter, 0.2 microns, ø 12.0 mm Use with: HMM100, HM34 Group: A 0195 Brass, sintered filter, 133 microns, ø 12.0 mm Use with: HMM100, HMP46 Group: A 10159HM 0195 6221 Plastic grid, 2 mm protection, ø 12.0 mm Use with: HM34, HMM100 Group: A 6221 230727SP Membrane filter, 0.2 microns, ø 12.
230602 Tube filter Use with: DMT132 Group: B 230602 231865 231865 Vaporized Hydrogen Peroxide resistant sintered filter Use with: HMT330, HMT360, HMT310, HM70, PTU300 Group: B Group C: 12 mm Diameter, Male Threads DRW010525 Membrane filter, 0.