Sensor Status List
Table 13 Sensor Status List
Value Description Notes
0 Sensor is working properly (OK).
1 Not measured yet.
2 Interface is not initialized.
1
1. Value is available only for sensors with a serial interface.
3 Communication timeout has occurred.
1
4 Unknown data is received.
1
5 Communication is functioning, but the sensor reports
errors. Use sensor's own service interface to find out
the cause.
1
6 Sensor communication is paused because service
connection is opened.
1
7 Message sequence numbers are overlapping in the
Autotrac satellite transceiver interface.
1
8 ... 19 Not available.
20 Excitation failure is caused by overload in the
excitation output.
2
2. Value is available only for sensors with a conventional, that is, analog or
counter/frequency, interface.
21 The input voltage is out of range or the A/D conversion
has failed due to an internal error.
2
22 Sensor is disconnected or the connection cables are
broken.
2
23 Sensor output exceeds the min/max limits defined in
the Measurements view.
2
24 Change in sensor output has exceeded the maximum
step defined in the Measurements view.
2
25 An internal configuration error has occurred.
2
26 Error in reference measurement, which is usually
caused by damaged sensor/logger or electrical
interference.
2
27 Internal voltage error occurred or the logger is
damaged.
2
28 PMT16 calibration data error.
2
29 Data is invalid for unspecified reason.
2
30 The measurement or the sensor has been manually
disabled.
99 Sensor status is not supported.