4.  From the list of parameters, select T
I/II
5.  At this point you need to select the adjustment method. If the 
reference probe is connected, two options are displayed: To 
same as T
I/II
, 1-point adjustment or 2-point adjustment. 
6.  You are prompted to check the settings, make sure that the 
pressure units of DM70 and DMT340 are the same. If not, 
change the unit by pressing   Unit on the corresponding row. 
To Same as T
I/II
NOTE 
This adjustment method automatically uses the reading of the 
reference probe as the reference value for making the adjustment. 
7.  After selecting To same as T
I/II
, confirm the adjustment by 
pressing   Yes. 
If you press   No, a note is displayed on the screen reminding 
you to follow the adjustment procedure described in the User's 
Guide. Press   OK and the adjustment mode display returns 
and no changes are made. 
1-Point Adjustment 
7. 
After selecting 1-point adjustment, press   Select to 
continue. 
8.  Set the probe to the reference temperature. You can follow the 
stabilization from the graph display by pressing 
 Graph. 
When the reading has stabilized, press 
 Ready. 
9.  Use the arrow buttons to give the reference temperature value 
and then press   OK. 
10.  Confirm the adjustment by pressing   Yes. (If you press   
No, the adjustment mode display returns and no changes are 
made.) The adjustment has now been completed. 
11.  To exit the adjustment mode, press   Back. 
12.  To return to the basic display, press   Exit. 