Adjusting Temperature 
Temperature Adjustment of DM70
The temperature adjustment can be done using the one-point or two-
point adjustment. The one-point adjustment should be done within the 
temperature range in which the device is most often used. 
To go to the adjustment mode, do the following: 
1. Press the Calibration button of the probe handle (see number 4 
in Figure 1 on page 11) using a tool with a thin, sharp point, for 
example, a small screwdriver. The indicator will switch into the 
adjustment mode. To confirm calibration, press   OK. 
2. Select T
d/f
 and press   Select. 
3. Select T and press   Select. 
4.  To select the adjustment method, first press   Adjust and then 
the adjustment method, either 1-point adjustment or 2-point 
adjustment. 
5.  Depending on your selection, continue from section One-Point 
Adjustment below or Two-Point Adjustment on page 75. 
One-Point Adjustment 
1. After selecting 1-point adjustment, press   Select to 
continue. 
2.  Set the probe to the reference temperature. You can follow the 
stabilization from the graph display by pressing 
 Graph. 
When the reading has stabilized, press   Ready. 
3.  Use the arrow buttons to give the reference temperature value 
and then press   OK. 
4.  Confirm the adjustment by pressing 
 Yes. (If you press   
No, the adjustment mode display returns and no changes are 
made.) The adjustment has now been completed. 
5.  To exit the adjustment mode, press 
 Back. 
6.  To return to the basic display, press   Exit. 