User`s guide
Chapter 9 _______________________________________ Calibration, Adjustment, and Maintenance 
17.  Place the screw back onto the calibration button. 
18.  To seal the calibration, attach a sticker on the calibration button. 
NOTE 
Several AutoCals can be necessary after this adjustment has been 
performed until the transmitter reaches full accuracy. 
1-Point Adjustment 
NOTE 
This adjustment method allows the customer to manually enter the 
reference value when making the adjustment. 
9.  A note is displayed on the screen reminding you to follow the 
adjustment procedure described in the User's Guide. Press   
OK to proceed with the adjustment. 
10.  Follow the stabilization from the graphic display by pressing   
Graph. The minimum recommended stabilization time for this 
critical adjustment is 5 hours. 
11.  When the reading has stabilized at the reference level, press   
Ready. 
12.  Use the arrow buttons to give the reference value. Then press   
OK. 
13.  Confirm the adjustment by pressing   Yes. If you press   No, 
the adjustment mode display returns and no changes are made. 
14.  Wait until the automatic auto-calibration has taken place. 
15.  When the calibration is completed, press   Back to exit the 
adjustment mode. 
16.  To return to the basic display, press   Exit. 
17.  Place the screw back onto the calibration button. 
18.  To seal the calibration, attach a sticker on the calibration button. 
NOTE 
Several AutoCals can be necessary after this adjustment has been 
performed until the transmitter reaches full accuracy. 
VAISALA _______________________________________________________________________ 73 