User`s guide
Chapter 9 _______________________________________ Calibration, Adjustment, and Maintenance 
Adjusting Dewpoint 
Before adjusting dewpoint, first carry out the two-point, relative 
humidity calibration, which will ensure the basic adjustment level. 
When adjusting the DMP74B or DMP74C probe or DMT340 series 
transmitters with the M-sensor, do the sensor purge before calibration 
and adjustment. For details, see section Sensor Purge on page 29. 
Two-Point Relative Humidity 
Adjustment 
NOTE 
For probes DMP74A, the wet point humidity reference must be >20 
%RH. 
For probes DMP74B and DMP74C, the wet point humidity reference 
must be <20 %RH. 
For DMP74A, the humidity references of 0 % (for example Nitrogen) 
and 30 ... 75 % are recommended. Make sure that the difference 
between the humidity references is more than 30 %. 
For DMP74B and DMP74C, the humidity references of 0 % (for 
example Nitrogen) and 11 % are recommended. 
To make the adjustment, follow the instructions below: 
1.  Check that DM70 is turned on. 
2.  When using the DMP74B probe, do the manual sensor purge. 
For details, see section Manual Sensor Purge on page 31. 
3.  Unscrew the screw from the probe handle to expose the 
calibration button (see num
ber 4 in Figure 1 on page 11). Then 
press the button with a sm
all screwdriver. When pressing the 
button, the indicator switches into the adjustment mode. 
4.  To start adjusting, press   OK. 
5. Then select RH and press   Select. 
6.  To check the environmental settings, press   Yes. To continue 
without checking the settings, press   No. 
VAISALA _______________________________________________________________________ 67 










