Chapter 9 _______________________________________ Calibration, Adjustment, and Maintenance 
CHAPTER 9 
CALIBRATION, ADJUSTMENT, AND 
MAINTENANCE 
Calibration 
DM70 is fully calibrated and adjusted as shipped from the factory. 
The typical calibration interval is one year. Calibration must always 
be done when there is reason to believe that the device is not within 
the accuracy specifications. 
The device can be sent to a Vaisala Service Center for calibration and 
adjustment, see contact information in section Vaisala Service Centers 
on page 78. 
User Calibration and Adjustment 
In this User's Guide the term "calibration" refers to comparing the 
device's reading against a reference value. "Adjustment" refers to 
changing the device's reading to correspond to the reference value. 
After the adjustment, the original calibration certificate shipped with 
the product is no longer valid. Read the instructions through carefully 
before making any adjustments. 
NOTE 
Dewpoint calibration should be carried out in Vaisala or in other 
laboratory conditions. 