User`s guide
Chapter 8 ____________________________Measuring Moisture in SFB6B Gas-insulated Equipment 
With DSC74B, the dewpoint can be measured at the GIE pressure. If 
the measurement is done in the GIE pressure, DM70 is able to show 
the corresponding dewpoint in the atmospheric pressure. 
DSC74B limits the flow rate with a leakage screw. The flow rate is 
optimized for GIE pressures between 3 ... 10 bar
g
, and no external 
flow meter is needed. The maximum flow can also be increased by 
removing the leak screw and adjusting the flow manually with the 
valve. To remove the leak screw, see Figure 30 below. 
  0405-012 
Figure 30  Removing the Leak Screw 
Parts per million by volume (ppm
v
) and parts per million by weight 
(ppm
w
) are frequently used to indicate humidity. The latter depends on 
the molecular weight of the gas. For SF
6
, the molecular weight is 
146.06, and the conversion between ppm
v
 and ppm
w
 is the following: 
ppm
w
 = ppm
v
 / 8.1 
DM70 is able to show the ppm
v
 and ppm
w
 values for moisture. 
VAISALA _______________________________________________________________________ 63 