User`s guide
Chapter 7 ____________________________________________________Sampling from Processes 
To make the connection, follow the instructions below: 
1.  Select the quick connector or thread adapter according to your 
process fitting. 
2.  Check that the pressure setting is correct. For details, see section 
Pressure Settings on page 36. 
3.  Make sure that the valve of the sam
pling cell is open. First close 
the valve, then turn it halfway open again. You can also first 
open the valve more to ventilate the parts, then readjust it to 
allow only a small leakage. 
4.  Carefully seal the threads of the quick connector or thread 
adapter with the PTFE thread seal tape. 
5.  Attach the quick connector or thread adapter onto the sample 
cell threads. Tighten the fitting with a fork spanner. 
6.  Connect the sample cell to the process fittings. Seal the fitting 
with the PTFE thread seal tape. 
7.  Install the gasket (delivered with the probe) to the nut of probe 
thread. 
VAISALA _______________________________________________________________________ 53 