User`s guide
USER'S GUIDE____________________________________________________________________ 
DMT242SC 
DMT242SC is a sampling cell with threads 3/8"G and 1/4"G. See 
Figure 24 on page 52. 
DMT242SC2 
DMT242SC2 is a sampling cell with 1/4" welded Swagelok 
connectors. See Figure 24 below. 
 0405-013 
Figure 24  DMT242SC and DMT242SC2 Sampling Cells 
Connection to Pressurized Processes Using 
the DSC74 Sampling Cell 
DSC74 is a Vaisala sampling cell for connecting DM70 to the 
pressurized processes (See section Accessories on page 89). With 
DSC74B and DSC74C m
easuring can be done in overpressure or 
atmospheric pressure. This depends on whether the sample gas is let 
into the sample cell before the needle valve or after that. See Figure 21 
on page 49 and Figure 22 on page 50. 
52 __________________________________________________________________ M010091EN-F 