User`s guide
Chapter 6 ____________________________________ Field Calibration Check of Fixed Transmitters 
1.  Connect the cable between DMP248 (connector X5, 
6-pin connector on the left of the motherboard) and DM70 
(connector I or II). Turn on both the devices. 
2.  Open the menu by pressing   Open. 
3. Select Functions and press  . 
4. Select Calibrate transmitters and press  . 
5. Select 1. Baud rate and 2.Serial format. To change the serial 
settings, press   Set. The serial settings of DMP248 and DM70 
must be the same. As a default, the DMP248 serial settings are 
4800 bps, 7 Even. 
6. Select 3. Start calibration and press 
 Start. If the serial settings of DM70 are not compatible with 
the settings of DM248, or the cable connection is not working, 
the following message appears: 
  0405-007 
Figure 19  Calibrate Transmitters Message 
Check that the serial settings of DMP248 and the settings of 
DM70 are the same. Check the cable connection. Then retry 
from step 6. 
7.  After the connection has been established, press 
 OK. 
8.  Select the quantity to check (T
d
 or T) and press 
 Select. 
9.  Check the environment settings of the DM70 probe, if needed. 
Press   Yes or   No. 
10.  The measurement value of DMP248 is shown on the upper row 
and the value of DM70 on the second row. The third row shows 
the difference between the measurements of DMP248 and 
DM70. 