USER'S GUIDE____________________________________________________________________ 
Turning off Automatic Auto-Calibration 
NOTE 
If the automatic auto-calibration is turned off, the auto-calibration 
should be started when starting measuring after the probe has not 
been used for a while and always at least once every hour. 
The automatic auto-calibration can be turned off as follows: 
1.  Open the menu by pressing 
 Open. 
2. Select 
Settings and press  . 
3. Select Measurement settings and press  . 
4.  To turn off the automatic auto-calibration, press 
 Off. To reactivate the automatic auto-calibration, press   
On. 
5.  To return to the basic display, press   Exit. 
Manual Auto-Calibration 
To check that auto-calibration has taken place for obtaining the most 
accurate measurement in a very dry environment, you can auto-
calibrate manually before measuring as follows: 
1.  Open the menu by pressing   Open. 
2. Select 
Functions and press  . 
3. Select AutoCal and press   Start. 
4.  If you want to start auto-calibration, press Yes. If the auto-
calibration conditions are not fulfilled, a note appears on the 
display informing that the calibration cannot be made or that a 
new attempt will be made later. 
5.  When the auto-calibration is completed, the display returns to 
the basic display automatically. 
Auto-calibration takes about 2 minutes. When the auto-calibration is 
in progress, an estimate of the measurement readings may appear on 
the display. Messages concerning the remaining auto-calibration time 
are shown periodically. See Figure 9 on page 29. 
NOTE 
If it is too humid, auto-calibration will not start. 