User`s guide
Chapter 5 _____________________________________________________________ User Interface 
4.  To turn rounding on (=a one-decimal display), press 
 On. To turn rounding off (=a two-decimal display), press   
Off. 
5.  To return to the basic display, press   Exit. 
NOTE 
Rounding does not affect the measurement accuracy. Accuracy is 
defined in Chapter 11 on page 81. 
Hold/Save 
With the Hold/Save function you can hold a selected display reading 
and save it into the memory. Several readings can be saved. The 
reading saved first is called data point 1, the second saved reading is 
called data point 2 , and so on. All the individual readings (data 
points) are stored in the same file marked with 
. The file will remain 
in the indicator memory even when the indicator is switched off. 
1.  Open the menu by pressing   Open. 
2. Select Display and then press  . 
3. Select Hold/Save display. 
4.  To hold the display, press   Hold. The measurement data is 
displayed. 
5.  To save the reading, press   Save. To return to the basic 
display, press   Exit. 
6.  To view the saved readings, press 
Record in the basic 
display. Then select  View recorded data and press  . 
7.  Select the file marked with 
 and press  . Now you can view 
the saved data readings. To view the recording time stamps, 
press 
 Times. 
8.  To return to the basic display, press 
 Exit. 
Graphic History 
The graphical display shows the measurements as a curve. From the 
curve, you can examine the data trend and history of the last minutes. 
VAISALA _______________________________________________________________________ 23 