HMI41 Calibration Option
19116ZZ-U218-1.1 Appendix 1
1996-08-26 5
2.4 Customized current output selection mode (cI)
DISPLAY WHAT TO DO PRESS:
Select the customized current output scale
selection mode (cI)
(number up) or
(number down)
ENTER to acknowledge the
selection
set
RH
Select the RH channel’s current output
range
(4...20 mA, 0...20 mA)
(number up) or
(number down)
ENTER to acknowledge the
selection
RH
set
Lo
%
Set the lower limit of the humidity range
(number up) or
(number down)
ENTER to acknowledge the
selection
RH
set
Hi
%
Set the upper limit of the humidity range
(number up) or
(number down)
ENTER to acknowledge the
selection
set
T
Select the T channel’s current output
range
(4...20 mA, 0...20 mA)
(number up) or
(number down)
ENTER to acknowledge the
selection
set
Lo
T
C
°C
Set the lower limit of the temperature
range
(number up) or
(number down)
ENTER to acknowledge the
selection
set
Hi
T
C
°C
Set the upper limit of the temperature
range
(number up) or
(number down)
ENTER to acknowledge the
selection
press ON/OFF