2.3 Customized voltage output selection mode (cU)
DISPLAY WHAT TO DO PRESS:
Select the customized voltage output
scale selection mode (cU)
▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
RH
set
Select the jumper selected RH channel’s
voltage output range
(0...1V, 0...5V, 0...10V)
▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
RH
set
Lo
Set the lower limit of the RH channel’s
voltage range
▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
RH
set
Hi
Set the upper limit of the RH channel’s
voltage range
▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
RH
set
Lo
%
Set the lower limit of the humidity range
▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
RH
set
Hi
%
Set the upper limit of the humidity range
▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
T
set
Select the jumper selected T channel’s
voltage output range
(0...1V, 0...5V, 0...10V)
▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
set
Lo
T
Set the lower limit of the T channel’s
voltage range
▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
set
Hi
T
Set the upper limit of the T channel’s
voltage range
▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
set
Lo
T
T°C
°C
Set the lower limit of the temperature
range
▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
set
Hi
T
T°C
°C
Set the upper limit of the temperature
range
▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
press ON/OFF