Manual

HMI41 Calibration Option
Appendix 1 19116ZZ-U218en-1.1
4 1996-08-26
2.3 Customized voltage output selection mode (cU)
DISPLAY WHAT TO DO PRESS:
Select the customized voltage output
scale selection mode (cU)
(number up) or
(number down)
ENTER to acknowledge the
selection
RH
set
Select the jumper selected RH channel’s
voltage output range
(0...1V, 0...5V, 0...10V)
(number up) or
(number down)
ENTER to acknowledge the
selection
RH
set
Lo
Set the lower limit of the RH channel’s
voltage range
(number up) or
(number down)
ENTER to acknowledge the
selection
RH
set
Hi
Set the upper limit of the RH channel’s
voltage range
(number up) or
(number down)
ENTER to acknowledge the
selection
RH
set
Lo
%
Set the lower limit of the humidity range
(number up) or
(number down)
ENTER to acknowledge the
selection
RH
set
Hi
%
Set the upper limit of the humidity range
(number up) or
(number down)
ENTER to acknowledge the
selection
T
set
Select the jumper selected T channel’s
voltage output range
(0...1V, 0...5V, 0...10V)
(number up) or
(number down)
ENTER to acknowledge the
selection
set
Lo
T
Set the lower limit of the T channel’s
voltage range
(number up) or
(number down)
ENTER to acknowledge the
selection
set
Hi
T
Set the upper limit of the T channel’s
voltage range
(number up) or
(number down)
ENTER to acknowledge the
selection
set
Lo
T
C
°C
Set the lower limit of the temperature
range
(number up) or
(number down)
ENTER to acknowledge the
selection
set
Hi
T
C
°C
Set the upper limit of the temperature
range
(number up) or
(number down)
ENTER to acknowledge the
selection
press ON/OFF