HMI41 Calibration Option
19116ZZ-U218-1.1 Appendix 1
1996-08-26 3
2.2 Current output scale selection mode (I)
DISPLAY WHAT TO DO PRESS:
Select the current output scale selection
mode (I)
(number up) or
(number down)
ENTER to acknowledge the
selection
set
Select the output
(0...20 mA or 4...20 mA)
(number up) or
(number down)
ENTER to acknowledge the
selection
set
Lo
T
C
°C
Set the lower limit of the temperature
range
(number up) or
(number down)
ENTER to acknowledge the
selection
set
Hi
T
C
°C
Set the upper limit of the temperature
range
(number up) or
(number down)
ENTER to acknowledge the
selection
ON/OFF