Insert the transmitter probe to the lower
humidity.
RH %
Let the readings stabilize.
MODE
RH
set
%
Set the blinking RH reading to the correct
value (e.g. to the equilibrium RH of the
salt solution).
V (number up) or
W (number down)
ENTER
RH
Insert the transmitter probe to the higher
humidity.
RH %
Let the readings stabilize.
MODE
RH %
set
Set the blinking RH reading to the correct
value (e.g. to the equilibrium RH of the
salt solution).
(number up) or
(number down)
ENTER to conclude
RH
Calibration has been successful.
ON/OFF
(disconnect the cable)
RH
Calibration has not been successful.
Perform the calibration again.