Owner manual

User's Guide ______________________________________________________________________
30 __________________________________________________________________ M210777EN-B
Display: What to Do Press
RH
Calibration has been successful.
ON/OFF
(disconnect the cable)
RH
Calibration has not been
successful. Perform the
calibration again.
Two-Point Calibration with HMI41 as a
Reference Meter
Display: What to Do Press
RH
set min max hyst hold Lo bat Hi
Pabs Pa
g/m
T
Td °C°F
°C°F
%
HMI41 has been turned on.
Indication of the software version (if the
version is 1.02 or more).
bat
Hi
Indication of the battery voltage.
RH
Select the desired calibration mode.
(number up) or
(number down)
ENTER
RH
Insert the HMI41 probe and the
transmitter probe to the lower humidity.
RH %
Let the readings stabilize. Alternatively,
you can change the display to show the
difference in readings.
ENTER to go on or
HOLD to have the
difference on the display
RH %
Let the readings stabilize. If you prefer,
you can return to the previous display.
ENTER to go on or
HOLD to return to the
previous display