page 1 of 17 Technology for Vacuum Systems Instructions for use VAP 5 DVR 5 Vacuum gauges Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. manual-no.
page 2 of 17 Dear customer, Your VACUUBRAND vacuum gauge shall support you at your work for a long time without any trouble and with full load output. Thanks to our large practical experience we attained much information how you could add to an efficient application and to personal safety. Please read these instructions for use prior to the initial start-up of your controller.
page 3 of 17 Contents Safety information! .................................................................................................... 4 Technical data ............................................................................................................ 6 Use and operation ..................................................................................................... 7 Troubleshooting ......................................................................................................
page 4 of 17 Safety information! Remove all packing materials, remove the equipment from its packing-box, remove the protective covers and inspect the equipment. If the equipment is damaged, notify the supplier and the carrier in writing within three days; state the item number of the equipment together with the order number and the suppliers´ invoice number. Retain all packing materials for inspection. Do not use the equipment if it is damaged.
page 5 of 17 Ensure that maintenance is done only by suitably trained and supervised technicians. Ensure that the maintenance technician is familiar with the safety procedures which relate to the equipment processed by the vacuum system and that the equipment if necessary is approximately decontaminated before starting maintenance. Use only genuine spare parts and accessories. ☞ Otherwise safety and performance of the equipment as well as the electromagnetic compatibility of the equipment might be reduced.
page 6 of 17 Technical data Type VAP 5 DVR 5 M e a s u r i n g ra n g e 1 0 0 0 m b a r - 1 0 -3 m b a r 1100 mbar - 0.1 mbar Measur ing pr inciple ther mal conduction, dependent on gas type capacitive, independent of gas type +/- 10% (of indicated value, within 1 0 0 m b a r - 1 0 -2 m b a r range)
page 7 of 17 Use and operation Display after switching on together with display ”CAL” (internal check and adjustment routine) adjustment mode warning symbol selectable pressure units Rear side rating plate serial interface (only DVR 5) mains switch key-button for digital adjustment and electronic scale conversion (pressure unit) ☞ at the side of the housing male connector for mains cable thread M8 for stand rod female connector for connecting cable to the gauge head Documents are only to be us
page 8 of 17 Description VAP 5 The vacuum gauge VAP 5 relies on the fact that the thermal conductivity of the residual gas in the vacuum chamber is a measure for the gas pressure. The thermal conductivity of gases is proportional to the pressure within a certain range and is related to the molecular mass. The thermal conductivity of gases and vapours varies with their molecular mass. The gauge has been adjusted for air (nitrogen) at the factory.
page 9 of 17 Interface parameters Interface parameters at the COM port have to be set as follows (the interface parameters at the DVR 5 cannot be changed): Interface parameter Baud rate 9600 Data bits 8 Star t bits 1 Stop bits 1 Parity none Handshake none Via the interface the actual pressure value can be read out. Interface command "IN_PV_1" readout of pressure value Output "XXXX.X [unit]" "-1" resolution of 0.
page 10 of 17 Troubleshooting Fault Possible cause R e m e dy No display. Î Mains not plugged in, no power available? ) Inser t mains plug, check m a i n s f u s e s. Î Other causes (device defective)? ) Retur n the device to the factor y for repair. Î Ther mal over load, is a m b i e n t t e m p e ra t u r e t o o high? ) Make sure ventilation is a d e q u a t e. Î Other causes (device defective)? ) Retur n the device to the factor y for repair.
page 11 of 17 Instructions for repair with directions for repair and spare parts list is available on request. ☞ The instructions are for trained service people. Cleaning the gauge head The gauge itself is maintenance-free. ➨ Only VSP 5: Using small screw driver or pencil tip, release locking ring in gauge head connection and remove protective filter with shim, keep locking ring. ➨ Fill the gauge head chamber with a solvent (e. g. benzene) and allow sufficient cleaning time.
page 12 of 17 Adjustment under vacuum: ➨ Evacuate gauge head to a pressure < 5 x 10-4 mbar. Status indication vacuum is displayed. Note: Adjustment under vacuum with an actual pressure > 5 x 10-4 mbar reduces the accuracy of the measurement. ➨ Push key-button at the side of the housing with a pencil tip or a small screwdriver. Status indication readjustment is displayed. ➨ Push key-button again in order to adjust pressure reading to 1 x 10-3 mbar. Status indication readjustment vanishes.
page 13 of 17 Adjustment under vacuum: ➨ Evacuate the device to a pressure < 5 x 10-2 mbar. Note: Adjustment under vacuum with an actual pressure higher than 5 x 10-2 mbar reduces the accuracy of measurement. If the pressure is significantly higher than 5 x 10-2 mbar, adjustment to a reference pressure is recommended. ➨ Push key-button at the side of the housing by using a pencil tip or a small screwdriver. The reading is automatically adjusted to “zero“.
page 14 of 17 Notes on return to the factory Repair - return - DKD calibration Safety and health of our staff, laws and regulations regarding the handling of dangerous goods, occupational health and safety regulations and regulations regarding safe disposal of waste require that for all pumps and other products the “Health and safety clearance form“ must be send to our office duly completed and signed before any equipment is dispatched to our premises.
page 15 of 17 Health and safety clearance form Declaration concerning safety, potential hazards and safe disposal of waste, e. g. used oil. Safety and health of our staff, laws and regulations regarding the handling of dangerous goods, occupational health and safety regulations, safety at work laws and regulations regarding safe disposal of waste, e. g.
page 16 of 17 Konformitätserklärung Declaration of conformity Déclaration de conformité Vakuum-Controller / Vacuum controller / Régulateur de vide DVR 5 (230V; 682910, 682911, 682912) VAP 5 (230V; 682850, 682858, 682881, 682880) Hiermit erklären wir, dass das oben bezeichnete Gerät in Konzeption und Bauart sowie in der von uns in Verkehr gebrachten Ausführung den grundlegenden Anforderungen der zutreffenden, aufgeführten EURichtlinien entspricht.
page 17 of 17 Disclaimer: Our technical literature is only intended to inform our customer. The validity of general empirical values and results obtained under test conditions for specific applications depend on a number of factors beyond our control. It is therefore strictly the users´ responsibility to very carefully check the validity of application to their specific requirements. No claims arising from the information provided in this literature will, consequently, be entertained.