Fault Possible cause Remedy
No display.
Power supply not
plugged in (wall plug)?
Plug in power supply.
CVC 3000 detect con-
troller switched off?
Switch on controller.
VACUU • BUS cable
of power supply not
plugged in at control-
ler?
Plug in VACUU • BUS
cable at controller.
Other than above men-
tioned causes?
Contact local distributor.
Display disappears.
Too much load (e.g.,
valves) connected?
Check current draw of the
connected devices.
Short circuit at connect-
ed valves?
Replace valves.
Short circuit at the
RS 232 plug?
Check plug and cable.
Other than above men-
tioned causes?
Contact local distributor.
Pressure reading
incorrect.
Pressure transducer
decalibrated?
Readjust CVC 3000 detect
or external gauge head.
Humidity in the pres-
sure transducer?
Let the pressure trans-
ducer dry, e.g., by pump-
ing. Readjust if neces-
sary. Determine and
eliminate the cause for
humidity.
Pressure transducer
contaminated?
See ”Cleaning the pres-
sure transducer”.
Other than above men-
tioned causes?
Contact local distributor.
Digital pressure
reading is ash-
ing, display shows
”0.0”.
Pressure transducer
not correctly adjusted
under vacuum?
Adjust CVC 3000 detect
or external gauge head
correctly.
No digital pressure
reading.
Pressure transducer
defective?
Contact local distributor.
Digital pressure
reading is ashing,
one blip*.
Overpressure at the
pressure transducer,
pressure > 795 Torr
(1060 mbar)?
Release pressure imme-
diately (risk of bursting).
Troubleshooting