Readjustment of the controller CVC 3000 detect
See section “Readjustment”, pg. 51.
Drain the solvent and dispose of in accordance with regulations. Re-
peat cleaning if necessary.
Rinse the measurement chamber several times with alcohol in order
to remove all solvent residues.
Allow the pressure transducer to dry.
Readjust the pressure transducer if necessary.