Assemble and lock the suitable power plug (included in
shipment) to the power supply prior to use.
The controller is equipped with a short-circuit-proof, wide-
range power supply with an integrated overload protec-
tion.
Position the controller and its vacuum line in such a way
that condensate cannot ow towards the pressure trans-
ducer.
In case connect inert gas at the venting connection.
Allow the equipment to equilibrate to ambient temperature
if you bring it from cold environment into a room prior to
operation. Notice if there is water condensation on cold
surfaces.
Comply with all applicable and relevant safety require-
ments (regulations and guidelines). Implement the re-
quired actions and adopt suitable safety measures.
Ambient conditions
Do not reach for this product if it has fallen into liquid.
There is a risk of deadly electrical shock. Unplug the
system immediately.
+ Do not use this product in an area where it can fall or
be pulled into water or other liquids.
Adopt suitable measures in case of differences from
recommended conditions, e.g., using the equipment
outdoors, conductive pollution or external condensa-
tion.
Do not operate this product near ames.
To the best of our knowledge the equipment is in com-
pliance with the requirements of the applicable EC-direc-
tives and harmonized standards (see ”Declaration of Con-
formity”) with regard to design, type and model. Directive
NOTICE
NOTICE